Development of gas chromatography-flame ionization detection system with a single column and liquid nitrogen-free for measuring atmospheric C2-C12 hydrocarbons.

作者信息

Liu Chengtang, Mu Yujing, Zhang Chenglong, Zhang Zhibo, Zhang Yuanyuan, Liu Junfeng, Sheng Jiujiang, Quan Jiannong

机构信息

Research Center for Eco-Environmental Sciences, Chinese Academy of Sciences, Beijing 100085, China.

Research Center for Eco-Environmental Sciences, Chinese Academy of Sciences, Beijing 100085, China.

出版信息

J Chromatogr A. 2016 Jan 4;1427:134-41. doi: 10.1016/j.chroma.2015.11.060. Epub 2015 Nov 21.

Abstract

A liquid nitrogen-free GC-FID system equipped with a single column has been developed for measuring atmospheric C2-C12 hydrocarbons. The system is consisted of a cooling unit, a sampling unit and a separation unit. The cooling unit is used to meet the temperature needs of the sampling unit and the separation unit. The sampling unit includes a dehydration tube and an enrichment tube. No breakthrough of the hydrocarbons was detected when the temperature of the enrichment tube was kept at -90 °C and sampling volume was 400 mL. The separation unit is a small round oven attached on the cooling column. A single capillary column (OV-1, 30 m × 0.32 mm I.D.) was used to separate the hydrocarbons. An optimal program temperature (-60 ∼ 170 °C) of the oven was achieved to efficiently separate C2-C12 hydrocarbons. There were good linear correlations (R(2)=0.993-0.999) between the signals of the hydrocarbons and the enrichment amount of hydrocarbons, and the relative standard deviation (RSD) was less than 5%, and the method detection limits (MDLs) for the hydrocarbons were in the range of 0.02-0.10 ppbv for sampling volume of 400 mL. Field measurements were also conducted and more than 50 hydrocarbons from C2 to C12 were detected in Beijing city.

摘要

已开发出一种配备单柱的无液氮气相色谱 - 火焰离子化检测系统,用于测量大气中的C2 - C12烃类。该系统由冷却单元、采样单元和分离单元组成。冷却单元用于满足采样单元和分离单元的温度需求。采样单元包括脱水管和富集管。当富集管温度保持在 -90°C且采样体积为400 mL时,未检测到烃类穿透。分离单元是安装在冷却柱上的小型圆形加热炉。使用一根毛细管柱(OV - 1,30 m×0.32 mm内径)分离烃类。实现了加热炉的最佳程序温度(-60 ∼ 170°C),以有效分离C2 - C12烃类。烃类信号与烃类富集量之间存在良好的线性相关性(R(2)=0.993 - 0.999),相对标准偏差(RSD)小于5%,对于400 mL的采样体积,烃类的方法检测限(MDLs)在0.02 - 0.10 ppbv范围内。还进行了现场测量,并在北京检测到50多种C2至C12的烃类。