Chen Keping, Zheng Xiaolin, Dai Yan, Wang Hao, Tang Yue, Lan Tingyu, Zhang Jinping, Tian Yi, Zhang Baojie, Zhou Xiaohong, Bonner Matthew, Zhang Shu
State Key Laboratory of Cardiovascular Disease, Arrhythmia Center, Fuwai Hospital, National Center for Cardiovascular Diseases, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing 100037, China.
Department of Echocardiography, Fuwai Hospital, National Center for Cardiovascular Diseases, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.
Europace. 2016 Nov;18(11):1748-1752. doi: 10.1093/europace/euv418. Epub 2016 Jan 31.
The miniaturized leadless pacemaker has recently emerged as a bradycardia therapy in humans, and many patients may need at least two pacemakers in their lifetime. Thus, the present study assessed the effect of two leadless pacemakers in the right ventricle (RV) on cardiac function in a swine model.
Fourteen mini pigs were chronically studied. Each animal received two sequential leadless pacemakers in the RV with 1 month between two separate implantations. All animals were then followed for 6 months. Cardiac function was assessed with echocardiography at baseline, the second implantation, and the end of the 6-month follow-up. A necropsy at the end of the study was performed to measure the length of the fibrous tissue covering the devices and assess tricuspid valve integrity. Four animals died during the surgical procedure, and one animal died of infection during the follow-up period. In the remaining nine animals, there were no significant changes in left ventricular ejection fraction, aortic time integral, cardiac output, and left ventricular size from baseline to the end of the study. The mean length of fibrous tissue covering Micra devices was 14.3 ± 7.8 mm. No tricuspid valve injury was observed.
Multiple leadless pacemakers implanted in the RV do not appear to impact cardiac function. Therefore, it appears safe to implant multiple leadless pacemakers in the same heart.
小型无导线起搏器最近已成为人类心动过缓的一种治疗方法,许多患者一生中可能至少需要两个起搏器。因此,本研究评估了在猪模型中右心室植入两个无导线起搏器对心脏功能的影响。
对14只小型猪进行了长期研究。每只动物在右心室先后植入两个无导线起搏器,两次单独植入之间间隔1个月。然后对所有动物进行6个月的随访。在基线、第二次植入时以及6个月随访结束时,用超声心动图评估心脏功能。在研究结束时进行尸检,以测量覆盖装置的纤维组织长度并评估三尖瓣完整性。4只动物在手术过程中死亡,1只动物在随访期间死于感染。在其余9只动物中,从基线到研究结束,左心室射血分数、主动脉时间积分、心输出量和左心室大小均无显著变化。覆盖Micra装置的纤维组织平均长度为14.3±7.8毫米。未观察到三尖瓣损伤。
右心室植入多个无导线起搏器似乎不会影响心脏功能。因此,在同一心脏中植入多个无导线起搏器似乎是安全的。