Automated compared to manual office blood pressure and to home blood pressure in hypertensive patients.

作者信息

Filipovský Jan, Seidlerová Jitka, Kratochvíl Zdeněk, Karnosová Petra, Hronová Markéta, Mayer Otto

机构信息

a Department of Internal Medicine II ;

b Biomedical Centre, Charles University , Medical Faculty in Pilsen , Pilsen , Czech Republic.

出版信息

Blood Press. 2016 Aug;25(4):228-34. doi: 10.3109/08037051.2015.1134086. Epub 2016 Feb 7.

Abstract

We studied the relationships of automated blood pressure (BP), measured in the healthcare centre, with manual office BP and home BP. Stable outpatients treated for hypertension were measured automatically, seated alone in a quiet room, six times after a 5 min rest with the BpTRU device, and immediately afterwards using the auscultatory method. Home BP was measured in a subgroup during 7 days preceding the visit. The automated, office and home BP values were 131.2 ± 21.8/77.8 ± 12.1 mmHg, 146.9 ± 20.8/85.8 ± 12.4 mmHg and 137.7 ± 17.7/79.4 ± 8.2 mmHg, respectively. Limits of agreement between office and automated BP (2 SDs in Bland-Altman plots) were +42.6 to -12.6/+22.6 to -6.6 mmHg for systolic/diastolic BP; for home and automated BP they were +45.8 to -25.8/+20.8 to -12.6 mmHg. For patients with two visits, intraclass correlation coefficients of BP values measured during the first and second visits were 0.66/0.72 for systolic/diastolic automated BP and 0.68/0.74 for systolic/diastolic office BP. Automated BP was lower than home BP and no more closely related to home BP than to office BP. It did not show better repeatability than office BP. Whether automated BP and the "white-coat effect", calculated cas the office BP-automated BP difference, have clinical and prognostic importance deserves further studies.

摘要

我们研究了在医疗中心测量的自动血压与诊室手动血压和家庭血压之间的关系。对接受高血压治疗的稳定门诊患者进行自动血压测量,患者独自坐在安静的房间里,使用BpTRU设备在休息5分钟后测量6次,随后立即采用听诊法测量。在就诊前7天对一个亚组患者测量家庭血压。自动血压、诊室血压和家庭血压值分别为131.2±21.8/77.8±12.1 mmHg、146.9±20.8/85.8±12.4 mmHg和137.7±17.7/79.4±8.2 mmHg。诊室血压与自动血压之间的一致性界限(Bland-Altman图中的2个标准差),收缩压/舒张压分别为+42.6至-12.6/+22.6至-6.6 mmHg;家庭血压与自动血压之间的一致性界限为+45.8至-25.8/+20.8至-12.6 mmHg。对于进行了两次就诊的患者,第一次和第二次就诊期间测量的血压值的组内相关系数,收缩压/舒张压自动血压分别为0.66/0.72,收缩压/舒张压诊室血压分别为0.68/0.74。自动血压低于家庭血压,与家庭血压的相关性并不比与诊室血压的相关性更密切。它的重复性并不比诊室血压更好。自动血压与“白大衣效应”(计算为诊室血压与自动血压之差)是否具有临床和预后重要性值得进一步研究。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验