Effectiveness of Anti-Calcification Technologies in a Rabbit Model.

作者信息

Wright Greg, de la Fuente Angela

出版信息

J Heart Valve Dis. 2015 May;24(3):386-92.

Abstract

BACKGROUND AND AIM OF STUDY

A kinetic profile revealing calcification progression is a powerful method for assessing the anti-calcification efficacy of various tissue treatments compared to the traditional analysis of a single time point in studies with rabbits.

METHODS

Tissue calcification was compared between test and control discs in rabbits. Test groups received either ThermaFix-processed bovine pericardium, or porcine leaflets processed with Linx, or alpha-amino oleic acid (AOA). Control groups received glutaraldehyde-treated porcine valve leaflets, and bovine pericardium. Tissue discs were implanted intramuscularly in rabbits and explanted every five days until day 75. Levels of calcium and phosphorus were then monitored in the rabbit tissues.

RESULTS

Calcium concentrations were plotted over time, with a sigmoidal curve being used for each kinetic profile. The ThermaFix group exhibited a lower calcium plateau level compared to the bovine control (107 ± 24 versus 258 ± 120 μg Ca2+ per mg dry tissue, respectively), while the Linx group exhibited a lower calcium plateau compared to porcine controls (130 ± 34 μg versus 280 ± 80 μg Ca2+ per mg dry tissue, respectively). The AOA group showed a low level of calcification through 40 days, but this was subsequently increased to a plateau of 258 ± 135 μg Ca2+ per mg dry tissue.

CONCLUSION

The study results showed that effects on the rate and level of calcification of anti-calcification technologies can be modulated. In particular, time is important when assessing tissue technology effectiveness.

摘要

研究背景与目的

与传统的在兔研究中对单个时间点进行分析相比,揭示钙化进展的动力学曲线是评估各种组织处理抗钙化效果的有力方法。

方法

比较兔试验组和对照组组织的钙化情况。试验组分别接受热固定处理的牛心包、用林克斯处理的猪瓣膜叶或α-氨基油酸(AOA)。对照组接受戊二醛处理的猪瓣膜叶和牛心包。将组织盘植入兔肌肉内,每5天取出直至第75天。然后监测兔组织中的钙和磷水平。

结果

绘制钙浓度随时间的变化曲线,每个动力学曲线均采用S形曲线。与牛对照组相比,热固定组的钙平台水平较低(分别为每毫克干组织107±24μg Ca2+和258±120μg Ca2+),而与猪对照组相比,林克斯组的钙平台较低(分别为每毫克干组织130±34μg和280±80μg Ca2+)。AOA组在40天内钙化水平较低,但随后增加到每毫克干组织258±135μg Ca2+的平台期。

结论

研究结果表明,抗钙化技术对钙化速率和水平的影响可以得到调节。特别是在评估组织技术有效性时,时间很重要。