[Calculation of the surface radial dose distribution of radioactive biliary stent loaded with 125I seeds by treatment planning system].

作者信息

Yao L L, Lin L, Sun H T, Wang J J, Zhou F G, Yan X Q, Teng G J, Guo J H

机构信息

Department of Radiation Oncology, Peking University Third Hospital, Beijing 100191, China.

出版信息

Zhonghua Yi Xue Za Zhi. 2016 Mar 8;96(9):727-30. doi: 10.3760/cma.j.issn.0376-2491.2016.09.014.

Abstract

OBJECTIVE

To study the surface radial dose distribution of different length radioactive biliary stents in different activity of (125)I seeds by treatment planning system.

METHODS

After a radioactive biliary stent was positioned in measurement phantom, which were made of solid paraffin and polymethyl methacrylate, a CT scan was performed to get the stent images. The images were then transferred to the treatment planning system for planning. The maximum dose level slice nearest to the center of the stent was selected to calculate the surface radial dose distribution.

RESULTS

The length of the stents (F=3 189.160, P<0.01) and the activity of the (125)I seeds (F= 811.509, P<0.01) can both significantly affect the cumulative radial dose distribution of the radioactive stent. Radial cumulative dose dose (Gy), stent length (cm), (125)I seeds activity (mCi) and distance from the stent surface (cm) meet the regression equation: ln dose =2.565+ 0.208 length+ 1.502 activity-0.738 distance (F=4 929.279, P<0.05).

CONCLUSIONS

The choice of suitable activity of radioactive (125)I should be based on treatment purpose in combination with the length and diameter of lesion and also with reference to the dose table. The measurement results are with smaller uncertainty, which can provide reference for the clinical application of dosimetry.