自主式阴道内远程监测:一项前瞻性临床验证研究。

Self-operated endovaginal telemonitoring: a prospective, clinical validation study.

作者信息

Pereira Isabel, von Horn Kyra, Depenbusch Marion, Schultze-Mosgau Askan, Griesinger Georg

机构信息

Department of Gynecological Endocrinology and Reproductive Medicine, University Hospital of Schleswig-Holstein, Campus Luebeck, Luebeck, Germany.

Department of Gynecological Endocrinology and Reproductive Medicine, University Hospital of Schleswig-Holstein, Campus Luebeck, Luebeck, Germany.

出版信息

Fertil Steril. 2016 Aug;106(2):306-310.e1. doi: 10.1016/j.fertnstert.2016.03.043. Epub 2016 Apr 16.

Abstract

OBJECTIVE

To study the comparability of self-operated endovaginal telemonitoring (SOET) with conventional two-dimensional transvaginal sonography (2D-TVS) monitoring during assisted reproductive technology (ART) cycles.

DESIGN

Single center, observational, single-blinded cohort study.

SETTING

University-affiliated in vitro fertilization center.

PATIENT(S): A total of 60 women undergoing ART cycles.

INTERVENTION(S): Explanation, training, and use of SOET system, and measurements of follicular and endometrial diameter with SOET and 2D-TVS.

MAIN OUTCOME MEASURE(S): Correlation of the total number of follicles >10 mm measured by SOET versus conventional 2D-TVS.

RESULT(S): In 16 cases (26.7%) the images were judged unsuitable for analysis. In these excluded cases the body mass index (BMI) was statistically significantly higher (29.3 vs. 24.4 kg/m(2)). The total number of follicles >10 mm was highly similar comparing SOET with conventional 2D-TVS (r = 0.91). For the concordance of whether more than 19 follicles or more than 25 follicles >10 mm were present, we found agreement between the methods in 43 of 44 cases (κ = 0.88) and 43 of 44 cases (κ = 0.85), respectively. For concordance on predefined human chorionic gonadotropin administration criteria, agreement was found in 39 of 44 cases (κ = 0.734).

CONCLUSION(S): The incidence of SOET videos not suitable for analysis seems to be associated with higher BMI. Otherwise, SOET showed good agreement with conventional 2D-TVS both for follicles and endometrium measurements. More importantly we also found good concordance regarding the cutoffs relevant for clinical decisions.

摘要

目的

研究在辅助生殖技术(ART)周期中,自主式阴道内远程监测(SOET)与传统二维经阴道超声检查(2D-TVS)监测的可比性。

设计

单中心、观察性、单盲队列研究。

设置

大学附属医院体外受精中心。

患者

共有60名接受ART周期的女性。

干预措施

对SOET系统进行解释、培训和使用,并使用SOET和2D-TVS测量卵泡和子宫内膜直径。

主要观察指标

SOET测量的直径>10 mm的卵泡总数与传统2D-TVS测量结果的相关性。

结果

16例(26.7%)图像被判定不适于分析。在这些被排除的病例中,体重指数(BMI)在统计学上显著更高(29.3对24.4 kg/m²)。将SOET与传统2D-TVS相比,直径>10 mm的卵泡总数高度相似(r = 0.91)。对于是否存在超过19个卵泡或超过25个直径>10 mm的卵泡的一致性,我们分别在44例中的43例(κ = 0.88)和44例中的43例(κ = 0.85)中发现两种方法之间具有一致性。对于预定义的人绒毛膜促性腺激素给药标准的一致性,在44例中的39例(κ = 0.734)中发现具有一致性。

结论

SOET视频不适于分析的发生率似乎与较高的BMI相关。否则,SOET在卵泡和子宫内膜测量方面与传统2D-TVS显示出良好的一致性。更重要的是,我们还发现在与临床决策相关的临界值方面具有良好的一致性。