Chan K K, Lou S L, Huang H K
Department of Radiological Sciences, University of California, Los Angeles 90024.
Radiology. 1989 Jun;171(3):847-51. doi: 10.1148/radiology.171.3.2717763.
Compression algorithms based on full-frame discrete cosine transforms have achieved compression ratios as high as 10:1 to 20:1 with almost imperceptible image degradation, when applied to projection radiographs digitized with 2,048 X 2,048 X 8-bit matrices. Compared with such radiographs, images obtained with computed tomography (CT) and magnetic resonance (MR) are smaller in size, have lower signal-to-noise ratios, and, in the case of CT, have a larger dynamic range. These differences result in qualitatively different spectral properties. The authors studied the efficiency of the full-frame technique when applied to CT and MR images. They achieved excellent results, with compression ratios in the neighborhood of 5:1. The study was performed with the use of a hardware implementation of the authors' algorithm, which can compress a 512 X 512 X 12-bit image in less than 1.5 seconds.