Suppr超能文献

碳离子扫描束治疗中4DCT伪影的剂量学影响:肺部和肝脏治疗的最坏情况分析

Dosimetric impact of 4DCT artifact in carbon-ion scanning beam treatment: Worst case analysis in lung and liver treatments.

作者信息

Mori Shinichiro, Kumagai Motoki, Karube Masataka, Yamamoto Naoyoshi

机构信息

Research Center for Charged Particle Therapy, National Institute of Radiological Sciences, Japan.

Research Center for Charged Particle Therapy, National Institute of Radiological Sciences, Japan.

出版信息

Phys Med. 2016 Jun;32(6):787-94. doi: 10.1016/j.ejmp.2016.05.003. Epub 2016 May 13.

Abstract

INTRODUCTION

We evaluated the impact of 4DCT artifacts on carbon-ion pencil beam scanning dose distributions in lung and liver treatment.

METHODS & MATERIALS: 4DCT was performed in 20 liver and lung patients using area-detector CT (original 4DCT). 4DCT acquisition by multi-detector row CT was simulated using original 4DCT by selecting other phases randomly (plus/minus 20% phases). Since tumor position can move over the respiratory range in original 4DCT, mid-exhalation was set as reference phase. Total prescribed dose of 60Gy (RBE) was delivered to the clinical target volume (CTV). Reference dose distribution was calculated with the original CT, and actual dose distributions were calculated with treatment planning parameters optimized using the simulated CT (simulated dose). Dose distribution was calculated by substituting these parameters into the original CT.

RESULTS

For liver cases, CTV-D95 and CTV-Dmin values for the reference dose were 97.6±0.5% and 89.8±0.6% of prescribed dose, respectively. Values for the simulated dose were significantly degraded, to 88.6±14.0% and 46.3±26.7%, respectively. Dose assessment results for lung cases were 84.8±12.8% and 58.0±24.5% for the simulated dose, showing significant degradation over the reference dose of 95.1±1.5% and 87.0±2.2%, respectively.

CONCLUSIONS

4DCT image quality should be closely checked to minimize degradation of dose conformation due to 4DCT artifacts. Medical staff should pay particular attention to checking the quality of 4DCT images as a function of respiratory phase, because it is difficult to recognize 4DCT artifact on a single phase in some cases.

摘要

引言

我们评估了4DCT伪影对肺部和肝脏治疗中碳离子笔形束扫描剂量分布的影响。

方法与材料

使用面积探测器CT(原始4DCT)对20例肝脏和肺部患者进行了4DCT检查。通过随机选择其他相位(加减20%相位),利用原始4DCT模拟多排探测器CT采集4DCT。由于在原始4DCT中肿瘤位置可在呼吸范围内移动,因此将呼气中期设为参考相位。向临床靶区(CTV)给予60Gy(RBE)的总处方剂量。用原始CT计算参考剂量分布,并用模拟CT优化的治疗计划参数计算实际剂量分布(模拟剂量)。通过将这些参数代入原始CT来计算剂量分布。

结果

对于肝脏病例,参考剂量的CTV-D95和CTV-Dmin值分别为处方剂量的97.6±0.5%和89.8±0.6%。模拟剂量的值显著降低,分别降至88.6±14.0%和46.3±26.7%。肺部病例的模拟剂量剂量评估结果分别为84.8±12.8%和58.0±24.5%,与参考剂量分别为95.1±1.5%和87.0±2.2%相比有显著下降。

结论

应密切检查4DCT图像质量,以尽量减少4DCT伪影导致的剂量适形性下降。医务人员应特别注意检查作为呼吸相位函数的4DCT图像质量,因为在某些情况下很难在单个相位上识别4DCT伪影。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验