基于相移检测的声管测量数据解读。

Interpretation of sonotubometric data based on phase-shift detection.

作者信息

Amoako-Tuffour Yaw, Garland Philip, Bance Manohar

机构信息

Faculty of Medicine, Dalhousie University, 1459 Oxford Street, Halifax, Nova Scotia, B3H 4R2, Canada.

Department of Mechanical Engineering, University of New Brunswick, 15 Dineen Drive, P.O. Box 4400, Fredericton, New Brunswick, E3B 5A3, Canada.

出版信息

J Otolaryngol Head Neck Surg. 2016 Jun 10;45(1):37. doi: 10.1186/s40463-016-0151-5.

Abstract

BACKGROUND

Sonotubometry is a non-invasive means of assessing Eustachian tube (ET) function. Its interpretation remains a complex task with questionable results due to wide variation between trials. A study was conducted to ascertain whether the measurement of phase shift in sonotubometric signals would be a more reliable indicator of ET patency than fluctuating Sound Pressure Level (SPL).

METHODS

The ears of six healthy participants and two participants with patulous ET (PET) were probed with a 100 Hz signal. Five recordings of SPL were performed at the external auditory canal. Cross-correlation was performed among filtered SPL signals and among extracted phase shift waveforms. Peak coefficients were averaged to provide a measure of waveform similarity between trials.

RESULTS

Mean peak cross correlation coefficient for SPL signal measured 0.603 ± 0.057 Standard Error of Mean (SEM) whilst that for Phase-Shift signal measured 0.884 ± 0.027 (SEM). All normal participants demonstrated an observable phase change between the ear and nasal signal during swallowing indicating an acoustic impedance change during the event. For the PET patients tested, the phase measurements in ear and nasal signals follow one another reasonably closely, indicating little or no impedance change during swallowing. It is thought that this impedance change is indicative of opening of the ET in normal patients, and the lack impedance change indicates ET either remaining open or remaining closed throughout the swallow.

CONCLUSIONS

Experimental data suggest that phase-shift detection is a more consistent means of interpreting sonotubometric data than SPL analysis.

摘要

背景

声管测量法是评估咽鼓管(ET)功能的一种非侵入性方法。由于各试验之间差异很大,其解读仍然是一项复杂的任务,结果也存在疑问。本研究旨在确定声管测量信号中的相移测量是否比波动的声压级(SPL)更可靠地指示咽鼓管通畅情况。

方法

用100Hz信号探测6名健康参与者和2名咽鼓管异常开放(PET)参与者的耳朵。在外耳道进行5次声压级记录。对滤波后的声压级信号之间以及提取的相移波形之间进行互相关分析。将峰值系数进行平均,以提供各试验之间波形相似性的度量。

结果

声压级信号的平均峰值互相关系数为0.603±0.057平均标准误差(SEM),而相移信号的平均峰值互相关系数为0.884±0.027(SEM)。所有正常参与者在吞咽过程中耳朵和鼻腔信号之间都表现出可观察到的相位变化,表明在此过程中声阻抗发生了变化。对于接受测试的咽鼓管异常开放患者,耳朵和鼻腔信号中的相位测量彼此相当接近,表明吞咽过程中阻抗变化很小或没有变化。据认为,这种阻抗变化表明正常患者的咽鼓管开放,而缺乏阻抗变化表明咽鼓管在整个吞咽过程中要么保持开放要么保持关闭。

结论

实验数据表明,与声压级分析相比,相移检测是解读声管测量数据更一致的方法。