Kristen K, Krempien B, Schulz P, Singer R, Wörle M, Heimke G
Dtsch Zahnarztl Z. 1978 May;33(5):311-8.
Investigations were carried out on implants of Al2O3 ceramic as secondary single toothe replacement in the lower jaw of young Beagle test dogs. Two basically different forms with altogether 7 variations were tested: the first three types of implants had conical roots and a further 4 had cylindrical roots. Common to all 7 types was an implantation shoulder which was tested for its ability to conduct the masticatory force via the cortical substance to the mandibular bone. All implants are at least under indirect masticatory strain; type 80-40-027-00 was directly under strain: part of the implant, shaped like a premolar protrudes through the gum directly into the oral cavity. Preliminary results show that cylindrical root shapes stood the test well. Their length was of no great importance. Conical root shapes without special provision of retention are not suitable. Firm positioning of the implant during introduction is of prime importance. A cortically placed implantation shoulder is not suitable for the conduction of the masticatory pressure into the mandibular bone in dogs.
对氧化铝陶瓷种植体进行了研究,该种植体作为年轻比格犬下颌单颗牙齿二次置换的植入物。测试了两种基本不同形式共7种变体:前三种植体有锥形牙根,另外4种有圆柱形牙根。所有7种植体的共同之处在于有一个植入肩部,测试了其通过皮质骨将咀嚼力传导至下颌骨的能力。所有种植体至少承受间接咀嚼应力;80 - 40 - 027 - 00型种植体直接承受应力:种植体的一部分形状像前磨牙,直接穿过牙龈突入口腔。初步结果表明,圆柱形牙根形状的种植体经测试效果良好。其长度并非十分重要。没有特殊固位措施的锥形牙根形状不合适。植入过程中种植体的稳固定位至关重要。置于皮质骨的植入肩部不适用于将咀嚼压力传导至犬类的下颌骨。