Kubiak Robert J, Lee Nancy, Zhu Yuan, Franch William R, Levitskaya Sophia V, Krishnan Surekha R, Abraham Varghese, Akufongwe Peter F, Larkin Christopher J, White Wendy I
Drug Metabolism, Pharmacokinetics and Biological Safety Assessment, MedImmune LLC, One MedImmune Way, Gaithersburg, MD 20878, USA.
Analytical Biotechnology Sciences and Strategy, MedImmune LLC, One MedImmune Way, Gaithersburg, MD 20878, USA.
J Immunol Res. 2016;2016:1485615. doi: 10.1155/2016/1485615. Epub 2016 Jul 10.
Consistent performance of anti-drug antibody (ADA) assays through all stages of clinical development is critical for the assessment of immunogenicity and interpretation of PK, PD, safety, and efficacy. The electrochemiluminescent assays commonly employed for ADA measurement use drug conjugated with ruthenium and biotin to bind ADA in samples. Here we report an association between high nonspecific ADA responses in certain drug-naïve individuals and the storage buffer of the conjugated reagents used in a monoclonal antibody ADA assay. Ruthenylated reagents stored in phosphate-buffered saline (PBS) buffer had increased levels of aggregate and produced variable and high baseline responses in some subjects. Reagents stored in a histidine-sucrose buffer (HSB) had lower aggregate levels and produced low sample responses. In contrast to PBS, conjugated reagents formulated in HSB remained low in aggregate content and in sample response variability after 5 freeze/thaw cycles. A reagent monitoring control (RMC) serum was prepared for the real-time evaluation of conjugated reagent quality. Using appropriate buffers for storage of conjugated reagents together with RMCs capable of monitoring of reagent aggregation status can help ensure consistent, long-term performance of ADA methods.
在临床开发的各个阶段,抗药抗体(ADA)检测的一致性对于免疫原性评估以及药代动力学(PK)、药效学(PD)、安全性和疗效的解读至关重要。常用于ADA检测的电化学发光检测法使用与钌和生物素偶联的药物来结合样品中的ADA。在此,我们报告了某些未接触过药物的个体中高非特异性ADA反应与单克隆抗体ADA检测中使用的偶联试剂储存缓冲液之间的关联。储存在磷酸盐缓冲盐水(PBS)缓冲液中的钌化试剂聚集水平升高,并且在一些受试者中产生了可变的高基线反应。储存在组氨酸 - 蔗糖缓冲液(HSB)中的试剂聚集水平较低,且样品反应也较低。与PBS不同,在HSB中配制的偶联试剂在5次冻融循环后聚集含量和样品反应变异性仍保持较低水平。制备了试剂监测对照(RMC)血清用于实时评估偶联试剂质量。使用合适的缓冲液储存偶联试剂并结合能够监测试剂聚集状态的RMC,有助于确保ADA检测方法的长期一致性性能。