Carey C M, Vogel G L
American Dental Association Health Foundation, Paffenbarger Research Center, and National Institute of Standards and Technology, Gaithersburg, MD 20899-8546.
J Res Natl Inst Stand Technol. 2000 Apr 1;105(2):267-73. doi: 10.6028/jres.105.030. Print 2000 Mar-Apr.
The activity of calcium in plaque fluid is needed to calculate the saturation level of that fluid relative to the tooth mineral. One method to determine the calcium activity in very small plaque fluid samples is by micro ion-selective electrode (ISE). Two commercially available calcium ionophores, a neutral-carrier and a charged-carrier, were evaluated in micro ISEs and compared to a commercially available macro ISE using saliva as a model for plaque fluid. The neutral-carrier containing ISEs gave results consistent with those of the macro ISE. Calcium activity measurements made with micro ISEs that contained the neutral ion-carrier of whole plaque samples and plaque fluid samples obtained by centrifugation of whole plaque showed that the activities did not change due to centrifugation. Estimates of the saturation with respect to hydroxyapatite were made from these measurements. A simplified calculation method is presented to estimate the ion activity product (IAP) of the calcium-phosphate minerals. The method is based on the relative abundance of some of the possible calcium-binding species and a fixed ionic strength for plaque fluid. Calculations show that within a normal pH range for plaque fluid (5.0 to 7.5) the differences in the IAP calculations for hydroxyapatite using the simplified method are less than those estimated from propagation of uncertainty calculations.
计算菌斑液相对于牙齿矿物质的饱和度需要菌斑液中钙的活性。测定非常小的菌斑液样本中钙活性的一种方法是使用微离子选择电极(ISE)。在微ISE中评估了两种市售的钙离子载体,一种中性载体和一种带电载体,并将其与使用唾液作为菌斑液模型的市售宏ISE进行比较。含有中性载体的ISE给出的结果与宏ISE的结果一致。用含有全菌斑样本中性离子载体的微ISE和通过全菌斑离心获得的菌斑液样本进行的钙活性测量表明,活性不会因离心而改变。根据这些测量结果估算了相对于羟基磷灰石的饱和度。提出了一种简化的计算方法来估算磷酸钙矿物质的离子活度积(IAP)。该方法基于一些可能的钙结合物种的相对丰度和菌斑液的固定离子强度。计算表明,在菌斑液的正常pH范围(5.0至7.5)内,使用简化方法计算羟基磷灰石的IAP的差异小于根据不确定度传播计算估计的差异。