Ebrahim Mohammed, Hegde Sanjeet, Printz Beth, Abcede Mark, Proudfoot James A, Davis Christopher
Division of Cardiology, Department of Pediatrics, University of California San Diego, Rady Children's Hospital San Diego, 3020 Children's Way, MC 5004, San Diego, CA, 92123, US.
Clinical and Translational Research Institute, University of California San Diego, San Diego, CA, US.
Pediatr Cardiol. 2016 Dec;37(8):1453-1457. doi: 10.1007/s00246-016-1456-x. Epub 2016 Aug 25.
Noninvasive measurement of cardiac output (CO) and particularly stroke volume (SV) remain difficult but potentially valuable. These variables can be particularly challenging to measure in children with congenital heart disease (CHD). Impedance cardiography (IC) is a technique shown to be accurate in measuring SV in adults and in children with structurally normal hearts. The ease of use and rapidity of SV measurement using IC makes it potentially attractive for young patients with CHD. Advances in IC technology have led to more sophisticated signal-morphology IC (SMIC) devices that may further improve accuracy. We tested the accuracy of SMIC to measure SV in 21 subjects with CHD by comparing measurements with those from cardiac magnetic resonance (CMR) imaging. There was good agreement between SMIC and CMR in measurement of SV: mean difference = 1.7 ml (p = 0.47); r = 0.89. The agreement and correlation persisted when controlling for the differences in blood pressure and heart rate during the two testing methods. We conclude that SMIC is accurate at measuring SV and thus CO when compared to CMR in a variety of forms of CHD.
无创测量心输出量(CO),尤其是每搏输出量(SV)仍然具有挑战性,但可能具有重要价值。对于患有先天性心脏病(CHD)的儿童来说,测量这些变量可能特别困难。阻抗心动图(IC)是一种在测量成人和心脏结构正常的儿童的SV时被证明准确的技术。使用IC测量SV的简便性和快速性使其对患有CHD的年轻患者具有潜在吸引力。IC技术的进步带来了更先进的信号形态IC(SMIC)设备,可能会进一步提高准确性。我们通过将测量结果与心脏磁共振(CMR)成像的测量结果进行比较,测试了SMIC在21名CHD患者中测量SV的准确性。SMIC和CMR在SV测量方面具有良好的一致性:平均差异 = 1.7毫升(p = 0.47);r = 0.89。在控制两种测试方法期间血压和心率的差异时,一致性和相关性仍然存在。我们得出结论,与CMR相比,SMIC在各种形式的CHD中测量SV以及因此测量CO时是准确的。