Mohan Megha, Lee Sau, Guo Changning, Peri S Prasad, Doub William H
Division of Pharmaceutical Analysis, Food and Drug Administration, 645 S Newstead Avenue, Saint Louis, Missouri, 63110, USA.
Office of Pharmaceutical Quality, Immediate Office, Food and Drug Administration, 10903 New Hampshire Avenue, Silver Spring, Maryland, 20993, USA.
AAPS PharmSciTech. 2017 Jul;18(5):1585-1594. doi: 10.1208/s12249-016-0625-9. Epub 2016 Sep 13.
The full-resolution next generation impactor (NGI) and three abbreviated impactor systems were used to obtain the apparent aerodynamic particle size distribution (APSD) and other quality measures for marketed dry powder inhalers (DPIs) using the compendial method and efficient data analysis (EDA). APSD for the active pharmaceutical ingredient (API) in Spiriva Handihaler, Foradil Aerolizer, and Relenza Diskhaler was obtained using a full-resolution NGI at 39, 60, and 90 L/min, respectively. Two reduced NGI (rNGI) configurations, the filter-only configuration (rNGI-f) and the modified-cup configuration (rNGI-mc), and the fast-screening impactor (FSI) with appropriate inserts to provide a 5-μm cut size were evaluated. The fine particle dose (FPD) obtained using the FSI for Spiriva was statistically similar to that obtained using the full NGI. However, the FPD for both Foradil and Relenza obtained using the FSI was significantly different from that obtained using the full NGI. Despite this, no significant differences were observed for the fine particle fraction (FPF) obtained using the FSI relative to that obtained from the full NGI for any of the DPIs. The use of abbreviated impactor systems appears promising with good agreement observed with the full-resolution NGI, except for small differences observed for the rNGI-mc configuration. These small differences may be product- and/or flow rate-specific, and further evaluation will be required to resolve these differences.
使用全分辨率下一代撞击器(NGI)和三种简化撞击器系统,通过药典方法和高效数据分析(EDA),获取市售干粉吸入器(DPI)的表观空气动力学粒径分布(APSD)和其他质量指标。分别在39、60和90升/分钟的流速下,使用全分辨率NGI获得思力华Handihaler、奥克斯都保和瑞乐砂碟式吸入器中活性药物成分(API)的APSD。评估了两种简化的NGI(rNGI)配置,即仅过滤器配置(rNGI-f)和改良杯配置(rNGI-mc),以及配备合适插入物以提供5微米切割粒径的快速筛选撞击器(FSI)。使用FSI获得的思力华细颗粒剂量(FPD)与使用全分辨率NGI获得的结果在统计学上相似。然而,使用FSI获得的奥克斯都保和瑞乐砂的FPD与使用全分辨率NGI获得的结果有显著差异。尽管如此,对于任何一种DPI,使用FSI获得的细颗粒分数(FPF)与使用全分辨率NGI获得的结果相比,均未观察到显著差异。除了rNGI-mc配置观察到的小差异外,简化撞击器系统的使用似乎很有前景,与全分辨率NGI观察到的一致性良好。这些小差异可能是特定产品和/或流速相关的,需要进一步评估以解决这些差异。