腕关节断层合成中的剂量降低

Dose Reduction in Tomosynthesis of the Wrist.

作者信息

Becker Anton S, Martini Katharina, Higashigaito Kai, Guggenberger Roman, Andreisek Gustav, Frauenfelder Thomas

机构信息

1 Department of Radiology, Institute of Diagnostic and Interventional Radiology, University Hospital Zurich, University of Zurich, Raemistrasse 100, 8091 Zurich, Switzerland.

出版信息

AJR Am J Roentgenol. 2017 Jan;208(1):159-164. doi: 10.2214/AJR.16.16729. Epub 2016 Oct 20.

Abstract

OBJECTIVE

The purpose of this study was to quantitatively and qualitatively determine the impact of radiation dose reduction on the image noise and quality of tomosynthesis studies of the wrist.

MATERIALS AND METHODS

Imaging of six cadaver wrists was performed with tomosynthesis in anteroposterior position at a tube voltage of 60 kV and tube current of 80 mA and subsequently at 60 or 50 kV with different tube currents of 80, 40, or 32 mA. Dose-area products (DAP) were obtained from the electronically logged protocol. Image noise was measured with an ROI. Two independent and blinded readers evaluated all images. Interreader agreement was measured with a Cohen kappa. Readers assessed overall quality and delineation of soft tissue, cortical bone, and trabecular bone on a 4-point Likert scale.

RESULTS

The highest DAP (3.892 ± 0.432 Gy · cm) was recorded for images obtained with 60 kV and 80 mA; the lowest (0.857 ± 0.178 Gy · cm) was recorded for images obtained with 50 kV and 32 mA. Noise was highest when a combination of 50 kV and 32 mA (389 ± 26.6) was used and lowest when a combination of 60 kV and 80 mA (218 ± 12.3) was used. The amount of noise on images acquired using 60 kV and 80 mA was statistically significantly different from the amount measured on all other images (p < 0.0001). Interreader agreement was excellent (κ = 0.93). Delineation of anatomy and overall quality were scored best on images obtained with 60 kV and 80 mA and worst on images obtained with 50 kV and 32 mA. The difference in delineation and quality on images obtained using 50 kV and 40 mA was not statistically significantly different compared with images obtained using 60 kV and 80 mA.

CONCLUSION

Significant dose reduction for tomosynthesis of the wrist is possible while image quality and delineation of anatomic structures remain preserved.

摘要

目的

本研究的目的是定量和定性地确定辐射剂量降低对腕关节断层合成研究的图像噪声和质量的影响。

材料与方法

对6具尸体腕关节进行成像,在管电压60 kV、管电流80 mA的条件下进行前后位断层合成成像,随后在60 kV或50 kV以及80、40或32 mA的不同管电流下进行成像。剂量面积乘积(DAP)从电子记录的协议中获取。使用感兴趣区域(ROI)测量图像噪声。两名独立且不知情的阅片者对所有图像进行评估。阅片者间的一致性用Cohen卡方进行测量。阅片者在4分李克特量表上评估软组织、皮质骨和小梁骨的整体质量及轮廓清晰度。

结果

60 kV和80 mA条件下获得的图像记录的DAP最高(3.892±0.432 Gy·cm);50 kV和32 mA条件下获得的图像记录的DAP最低(0.857±0.178 Gy·cm)。使用50 kV和32 mA组合时噪声最高(389±26.6),使用60 kV和80 mA组合时噪声最低(218±12.3)。60 kV和80 mA条件下获取的图像上的噪声量与所有其他图像上测量的噪声量在统计学上有显著差异(p<0.0001)。阅片者间的一致性极佳(κ=0.93)。在60 kV和80 mA条件下获得的图像上,解剖结构轮廓清晰度和整体质量评分最佳,在50 kV和32 mA条件下获得的图像上评分最差。与60 kV和80 mA条件下获得的图像相比,50 kV和40 mA条件下获得的图像在轮廓清晰度和质量上的差异在统计学上无显著差异。

结论

在保留图像质量和解剖结构轮廓清晰度的同时,腕关节断层合成有可能大幅降低辐射剂量。