Salazar Antonio J, Romero Javier A, Bernal Oscar A, Moreno Angela P, Velasco Sofía C, Díaz Xavier A
Electrophysiology and Telemedicine Laboratory, University of Los Andes, Carrera 1 Este No. 19A-40, Bogotá 11001, Colombia.
Department of Diagnostic Imaging, Fundación Santa Fe de Bogotá University Hospital, Calle 119 No. 7-75, Bogotá 11001, Colombia; School of Medicine, University of Los Andes, Carrera 1 Este No. 19A-40, Bogotá 11001, Colombia.
Int J Telemed Appl. 2016;2016:3642960. doi: 10.1155/2016/3642960. Epub 2016 Sep 29.
The aim of this study was to evaluate and compare the clinical performance of different alternatives to implement low-cost screening telemammography. We compared computed radiography, film printed images, and digitized films produced with a specialized film digitizer and a digital camera. The ethics committee of our institution approved this study. We assessed the equivalence of the clinical performance of observers for cancer detection. The factorial design included 70 screening patients, four technological alternatives, and cases interpreted by seven radiologists, for a total of 1,960 observations. The variables evaluated were the positive predictive value (PPV), accuracy, sensitivity, specificity, and the area under the receiver operating characteristic curves (AUC). . The mean values for the observed variables were as follows: accuracy ranged from 0.77 to 0.82, the PPV ranged from 0.67 to 0.68, sensitivity ranged from 0.64 to 0.74, specificity ranged from 0.87 to 0.90, and the AUC ranged from 0.87 to 0.90. At a difference of 0.1 to claim equivalence, all alternatives were equivalent for all variables. Our findings suggest that telemammography screening programs may be provided to underserved populations at a low cost, using a film digitizer or a digital camera.
本研究的目的是评估和比较实施低成本筛查远程乳腺摄影不同方法的临床性能。我们比较了计算机X线摄影、胶片打印图像以及使用专用胶片数字化仪和数码相机生成的数字化胶片。我们机构的伦理委员会批准了本研究。我们评估了观察者在癌症检测方面临床性能的等效性。析因设计包括70名筛查患者、四种技术方法以及由七名放射科医生解读的病例,共计1960次观察。评估的变量包括阳性预测值(PPV)、准确性、敏感性、特异性以及受试者工作特征曲线下面积(AUC)。观察变量的平均值如下:准确性在0.77至0.82之间,PPV在0.67至0.68之间,敏感性在0.64至0.74之间,特异性在0.87至0.90之间,AUC在0.87至0.90之间。若差异为0.1时可宣称等效,则所有方法在所有变量上均等效。我们的研究结果表明,使用胶片数字化仪或数码相机,可低成本地为服务不足的人群提供远程乳腺摄影筛查项目。