Fowler Joel B
National Institute of Standards and Technology, Gaithersburg, MD 20899-0001.
J Res Natl Inst Stand Technol. 1996 Sep-Oct;101(5):629-637. doi: 10.6028/jres.101.062.
A high temperature oil-bath-based-black-body source has been designed and constructed in the Radiometric Physics Division at the National Institute of Standards and Technology, Gaithersburg, MD. The goal of this work was to design a large aperture blackbody source with highly uniform radiance across the aperture, good temporal stability, and good reproducibility. This blackbody source operates in the 293 K to 473 K range with blackbody temperature combined standard uncertainties of 7.2 mK to 30.9 mK. The calculated emissivity of this source is 0.9997 with a standard uncertainty of 0.0003. With a 50 mm limiting aperture at the cavity entrance, the emissivity increases to 0.99996.
美国国家标准与技术研究院辐射物理部(位于马里兰州盖瑟斯堡)设计并建造了一个基于高温油浴的黑体源。这项工作的目标是设计一个大孔径黑体源,其孔径上具有高度均匀的辐射率、良好的时间稳定性和良好的可重复性。该黑体源在293 K至473 K范围内运行,黑体温度的合成标准不确定度为7.2 mK至30.9 mK。该源的计算发射率为0.9997,标准不确定度为0.0003。在腔体入口处有一个50 mm的限制孔径时,发射率增加到0.99996。