无柄肩关节置换术中的透射线现象与一种影像学表现相关。

Radiolucency in stemless shoulder arthroplasty is associated with an imaging phenomenon.

作者信息

Hudek Robert, Werner Birgit, Abdelkawi Ayman F, Schmitt Rainer, Gohlke Frank

机构信息

RHÖN Klinikum AG, Klinik für Schulterchirurgie, Salzburger Leite 1, D-97616 Bad Neustadt, Germany.

RHÖN Klinikum AG, Klinik für Radiologie, Salzburger Leite 1, D-97616 Bad Neustadt, Germany.

出版信息

J Orthop Res. 2017 Sep;35(9):2040-2050. doi: 10.1002/jor.23478. Epub 2017 Apr 21.

Abstract

Stemless humeral implants show comparable midterm clinical results compared to stemmed components. Recently, radiolucencies around the metaphyseal seating of humeral stemless implants were reported on postoperative radiographs. It is controversial whether they are attributable to bone resorption. We hypothesized these radiolucencies result from imaging artifacts. Seven cadaveric specimens (three male and four female) were first radiographed and then scanned with CT. A stemless humeral component of current design was implanted in each specimen. After implantation, all specimens were radiographed with different exposure settings. The implant was removed, and the specimens were scanned with CT again. Pre- and post-implantation radiographs and CT scans were compared. The mean Hounsfield units (HU) at the humeral resection plane from the pre-implantation CT were correlated with the diameter of the radiolucent halo on the post-implantation radiographs. A symmetric radiolucent halo of variable diameters occurred on all radiographs after implantation when an automatic exposure control was used. The halo disappeared in all specimens when the tube voltage was reduced. Lower CT-values (HU) before the implantation resulted in greater halos on the radiograph after implantation. Symmetric radiolucent halos can result from imaging artifacts, which is most likely due to radiation scatter. The halos can be minimized by reducing the tube voltage. The halo effect appears to be pronounced in bones with decreased density. © 2017 Orthopaedic Research Society. Published by Wiley Periodicals, Inc. J Orthop Res 35:2040-2050, 2017.

摘要

与有柄组件相比,无柄肱骨植入物显示出相当的中期临床结果。最近,术后X线片报告了无柄肱骨植入物干骺端固定部位周围的透射线区。这些透射线区是否归因于骨吸收存在争议。我们推测这些透射线区是由成像伪影导致的。首先对7具尸体标本(3男4女)进行X线摄影,然后进行CT扫描。将当前设计的无柄肱骨组件植入每个标本中。植入后,所有标本用不同的曝光设置进行X线摄影。取出植入物,再次对标本进行CT扫描。比较植入前后的X线片和CT扫描结果。将植入前CT扫描肱骨切除平面处的平均亨氏单位(HU)与植入后X线片上透射线晕圈的直径进行相关性分析。当使用自动曝光控制时,植入后所有X线片上均出现了直径可变的对称透射线晕圈。当管电压降低时,所有标本中的晕圈均消失。植入前较低的CT值(HU)导致植入后X线片上的晕圈更大。对称透射线晕圈可能是由成像伪影导致的,最有可能是由于辐射散射。通过降低管电压可将晕圈最小化。在密度降低的骨骼中,晕圈效应似乎更为明显。©2017骨科学研究协会。由威利期刊公司出版。《矫形外科学研究》2017年第35卷:2040 - 2050页。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验