Understanding reconstructed Dante spectra using high resolution spectroscopy.

作者信息

May M J, Weaver J, Widmann K, Kemp G E, Thorn D, Colvin J D, Schneider M B, Moore A, Blue B E

机构信息

L-170 Lawrence Livermore National Laboratory, 7000 East Ave., Livermore, California 94551, USA.

Naval Research Laboratory, 4555 Overlook Ave. SW, Washington, DC 20375, USA.

出版信息

Rev Sci Instrum. 2016 Nov;87(11):11E330. doi: 10.1063/1.4961267.

Abstract

The Dante is an 18 channel filtered diode array used at the National Ignition Facility (NIF) to measure the spectrally and temporally resolved radiation flux between 50 eV and 20 keV from various targets. The absolute flux is determined from the radiometric calibration of the x-ray diodes, filters, and mirrors and a reconstruction algorithm applied to the recorded voltages from each channel. The reconstructed spectra are very low resolution with features consistent with the instrument response and are not necessarily consistent with the spectral emission features from the plasma. Errors may exist between the reconstructed spectra and the actual emission features due to assumptions in the algorithm. Recently, a high resolution convex crystal spectrometer, VIRGIL, has been installed at NIF with the same line of sight as the Dante. Spectra from L-shell Ag and Xe have been recorded by both VIRGIL and Dante. Comparisons of these two spectroscopic measurements yield insights into the accuracy of the Dante reconstructions.