Suppr超能文献

一种特殊设计的高强度线圈与标准圆形线圈经颅磁刺激装置的比较。

Comparison of a special designed high intensity coil to a standard round coil-TMS-setting.

作者信息

Szelényi Andrea, Wölfle Julia, Fischer Igor, Steiger Hans-Jakob, Sabel Michael

机构信息

Department of Neurosurgery, Hospital of the Ludwig Maximilians University Munich (LMU), Munich, Germany.

Department of Neurosurgery, University Hospital Duesseldorf, Medical Faculty, Heinrich-Heine-University, Duesseldorf, Germany.

出版信息

J Neurosci Methods. 2017 Mar 1;279:72-78. doi: 10.1016/j.jneumeth.2017.01.006. Epub 2017 Jan 16.

Abstract

BACKGROUND

Routine diagnostic Transcranial Magnetic Stimulation (TMS) is performed with a round coil (RC) for cranial and spinal root stimulation, being less successful for motor evoked potentials (MEP) of lower limb muscles. MEP elicited with a special configured flat figure-of-eight coil designed for high intensity stimulation (HI-coil) were compared to RC with regard to handling, efficiency, and physiological properties of MEP.

NEW METHODS

MEP elicited with HI-coil and 9-cm diameter RC for cortical, spinal and peripheral stimulation (PES) were compared for Motor threshold (MT), latencies and amplitudes of bilateral Abductor pollicis brevis (APB) and Abductor hallucis muscles (AH).

RESULTS AND COMPARISON WITH EXISTING METHODS

MT for HI-coil were significantly lower for cortical and spinal root stimulation compared to RC (APB: 37% vs. 48%; AH: 58% vs. 72%). MEP-latencies elicited with HI-coil and RC were without significant difference. AH-MEP amplitudes were significantly larger for HI-coil cortical (705±980μV vs. 370±280μV) and root stimulation (260±210μV vs. 151±100μV). Amplitudes elicited by PES compared to HI-coil were always significantly larger.

CONCLUSION

Results for cortical and spinal root stimulation with regards to latencies and amplitudes for APB were equivalent between HI-coil and RC. PES was superior in achieving supramaximal stimulation in terms of amplitudes. The use of the HI-coil might be advantageous for MEP of lower extremity muscles with emphasis on pathologic conditions requiring higher stimulation intensities.

摘要

背景

常规诊断性经颅磁刺激(TMS)使用圆形线圈(RC)进行颅脑和脊髓神经根刺激,但在下肢肌肉运动诱发电位(MEP)方面效果欠佳。将一种为高强度刺激设计的特殊配置扁平8字形线圈(HI线圈)引出的MEP,在操作、效率及MEP生理特性方面与RC进行比较。

新方法

比较HI线圈和直径9厘米的RC引出的用于皮质、脊髓和外周刺激(PES)的MEP的运动阈值(MT)、双侧拇短展肌(APB)和拇展肌(AH)的潜伏期及波幅。

结果及与现有方法的比较

与RC相比,HI线圈用于皮质和脊髓神经根刺激时MT显著更低(APB:37%对48%;AH:58%对72%)。HI线圈和RC引出的MEP潜伏期无显著差异。HI线圈用于皮质刺激时AH-MEP波幅显著更大(705±980μV对370±280μV),用于神经根刺激时也更大(260±210μV对151±100μV)。与HI线圈相比,PES引出的波幅总是显著更大。

结论

HI线圈和RC在APB潜伏期和波幅方面用于皮质和脊髓神经根刺激的结果相当。在波幅方面,PES在实现超强刺激方面更具优势。对于需要更高刺激强度的病理状况,使用HI线圈对下肢肌肉的MEP可能具有优势。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验