Cai Anji, Qi Suwen, Su Zhuowa, Yang Yu, Shen Huaqing, Wang Tianyuan, Dai Yong
Clin Lab. 2017 Jan 1;63(1):73-77. doi: 10.7754/Clin.Lab.2016.160536.
A comparison of any two methods is of great importance in a clinical laboratory. In this study, our aim is to compare the assay results of blood urea nitrogen (BUN), creatinine (Cr), and uric acid (UA) obtained through two distinct methods and then assess the analytical agreement of the two methods.
A test method (Vitros5600 system) measuring BUN, Cr, and UA analytes was compared with a reference method (Hitachi7600 system). The Clinical and Laboratory Standards Institute (CLSI) document EP9-A2 guidelines were followed to evaluate the method comparison and bias using 40 patient samples.
A high correlation between the two methods was found for all of the samples (R2 > 0.990). The regression parameters were BUN (R2 = 0.9996, slope = 1.025, intercept = 0.1156), Cr (R2 = 0.9993, slope = 0.9993, intercept = 4.661), and UA (R2 = 0.9971, slope = 1.011, intercept = 1.311). Compared with the Hitachi7600 reference method, the Vitros5600 test method showed that the 95% confidence interval for the predicted bias at medical decision levels was less than the acceptable error. More importantly, Bland-Altman plots indicated that a minimal positive bias (mean ± SD) was observed: BUN (0.352 ± 0.289 mmol/L), Cr (2.702 ± 7.683 µmol/L), UA (5.398 ± 7.086 µmol/L).
The Vitros5600 and Hitachi7600 systems have good correlation and bias for detecting BUN, Cr, and UA analytes. The two systems have a high method agreement.
在临床实验室中,比较任意两种方法非常重要。在本研究中,我们的目的是比较通过两种不同方法获得的血尿素氮(BUN)、肌酐(Cr)和尿酸(UA)的检测结果,然后评估这两种方法的分析一致性。
将测量BUN、Cr和UA分析物的测试方法(Vitros5600系统)与参考方法(日立7600系统)进行比较。遵循临床和实验室标准协会(CLSI)文件EP9 - A2指南,使用40份患者样本评估方法比较和偏差。
所有样本在两种方法之间均发现高度相关性(R2>0.990)。回归参数为BUN(R2 = 0.9996,斜率 = 1.025,截距 = 0.1156)、Cr(R2 = 0.9993,斜率 = 0.9993,截距 = 4.661)和UA(R2 = 0.9971,斜率 = 1.011,截距 = 1.311)。与日立7600参考方法相比,Vitros5600测试方法显示,在医学决定水平下预测偏差的95%置信区间小于可接受误差。更重要的是,Bland - Altman图表明观察到最小的正偏差(均值±标准差):BUN(0.352±0.289 mmol/L)、Cr(2.702±7.683 µmol/L)、UA(5.398±7.086 µmol/L)。
Vitros5600和日立7600系统在检测BUN、Cr和UA分析物方面具有良好的相关性和偏差。这两个系统具有高度的方法一致性。