Urli D, Calon B, Clever B, Pottecher T
Service d'anesthésie et de réanimation chirurgicale, Strasbourg.
Ann Fr Anesth Reanim. 1989;8(4):369-70. doi: 10.1016/s0750-7658(89)80082-6.
Lung overinflation was observed in a patient ventilated by a Siemens Servo Ventilator 900 C. The expiratory valves failure to open was related to a transducer disconnection in the expiratory limb. This transducer controls opening of the expiratory valve and when disconnected expiratory valve remains closed.
在一名使用西门子Servo Ventilator 900 C呼吸机进行通气的患者中观察到肺过度充气。呼气阀未能打开与呼气支路中的传感器断开有关。该传感器控制呼气阀的打开,断开连接时呼气阀保持关闭状态。