Abstract

Metal-induced encephalopathy after stent-assisted coil embolization is extremely rare. The present report describes two patients who presented with symptomatic intracranial parenchymal edematous lesions after stent-assisted coil embolization. A 64-year-old woman underwent stent-assisted coil embolization for a left internal carotid artery aneurysm; 21 days after the procedure she presented with right hand weakness and MRI revealed multifocal white matter lesions. Another woman aged 52 years underwent stent-assisted coil embolization for right vertebral artery aneurysm; 18 days after the procedure she presented with left-sided sensory disturbance and MRI demonstrated multiple white matter lesions. Treatment in both cases resulted in improvement of these lesions after steroid pulse therapy, and the patients had no associated morbidity 4 months after the procedures. Clinicians should monitor for neurologic symptoms and postoperative delayed radiologic parenchymal edematous changes associated with the metal allergic reaction after nitinol stent-assisted coil embolization.

摘要

支架辅助弹簧圈栓塞术后金属诱导的脑病极为罕见。本报告描述了两名患者,他们在支架辅助弹簧圈栓塞术后出现了有症状的颅内实质水肿性病变。一名64岁女性因左侧颈内动脉动脉瘤接受了支架辅助弹簧圈栓塞术;术后21天,她出现右手无力,MRI显示多灶性白质病变。另一名52岁女性因右侧椎动脉动脉瘤接受了支架辅助弹簧圈栓塞术;术后18天,她出现左侧感觉障碍,MRI显示多个白质病变。两例患者在接受类固醇脉冲治疗后,这些病变均有所改善,术后4个月患者无相关并发症。临床医生应监测镍钛合金支架辅助弹簧圈栓塞术后与金属过敏反应相关的神经症状和术后延迟的放射学实质水肿变化。

相似文献

2
Nickel-associated delayed multiple white matter lesions after stent-assisted coil embolization of intracranial unruptured aneurysm.
J Neurointerv Surg. 2018 Jan;10(1):e1. doi: 10.1136/neurintsurg-2017-013005.rep. Epub 2017 Feb 24.
3
Delayed Enhancing White Matter Lesions, a Rare Complication After Stent-Assisted Coil Embolization: A Case Report.
World Neurosurg. 2024 Jul;187:42-45. doi: 10.1016/j.wneu.2024.03.110. Epub 2024 Mar 26.
5
Delayed recurrent enhancing white matter lesions complicating coiling of intracranial aneurysm.
Eur J Neurol. 2021 Jul;28(7):2388-2391. doi: 10.1111/ene.14844. Epub 2021 Apr 7.
6
Delayed coil migration from a small wide-necked aneurysm after stent-assisted embolization: case report and literature review.
Neuroradiology. 2006 May;48(5):333-7. doi: 10.1007/s00234-005-0044-1. Epub 2006 Apr 6.
10
Incidence and risk factors of postoperative headache after endovascular coil embolization of unruptured intracranial aneurysms.
Acta Neurochir (Wien). 2014 Jul;156(7):1281-7. doi: 10.1007/s00701-014-2095-8. Epub 2014 May 7.

引用本文的文献

1
Stent-Assisted Coiling in a Nickel-Allergic Patient.
Cureus. 2025 Apr 27;17(4):e83055. doi: 10.7759/cureus.83055. eCollection 2025 Apr.
2
Reversible and multiphasic parenchymal changes in MRI after coil embolization for a ruptured cerebral aneurysm.
Surg Neurol Int. 2023 Apr 21;14:147. doi: 10.25259/SNI_160_2023. eCollection 2023.
4
Delayed Leukoencephalopathy: A Rare Complication after Coiling of Cerebral Aneurysms.
AJNR Am J Neuroradiol. 2020 Feb;41(2):286-292. doi: 10.3174/ajnr.A6386. Epub 2020 Jan 30.

本文引用的文献

1
Safety and Efficacy of Wingspan Stenting for Severe Symptomatic Atherosclerotic Stenosis of the Middle Cerebral Artery: Analysis of 278 Continuous Cases.
J Stroke Cerebrovasc Dis. 2016 Oct;25(10):2368-72. doi: 10.1016/j.jstrokecerebrovasdis.2016.05.035. Epub 2016 Jun 17.
2
Cell-mediated allergy to cerebral aneurysm clip causing extensive cerebral edema.
J Neurosurg. 2014 Oct;121(4):924-8. doi: 10.3171/2014.6.JNS132405. Epub 2014 Jul 4.
3
North American Contact Dermatitis Group patch test results: 2009 to 2010.
Dermatitis. 2013 Mar-Apr;24(2):50-9. doi: 10.1097/DER.0b013e3182819c51.
5
Hypersensitivity reactions to metallic implants - diagnostic algorithm and suggested patch test series for clinical use.
Contact Dermatitis. 2012 Jan;66(1):4-19. doi: 10.1111/j.1600-0536.2011.01971.x. Epub 2011 Sep 29.
7
Nickel release after implantation of the Amplatzer occluder.
Am Heart J. 2003 Apr;145(4):737-41. doi: 10.1067/mhj.2003.7.
8
Ability of Ni-containing biomedical alloys to activate monocytes and endothelial cells in vitro.
J Biomed Mater Res. 1999 Jun 5;45(3):251-7. doi: 10.1002/(sici)1097-4636(19990605)45:3<251::aid-jbm13>3.0.co;2-5.
9
Cell-mediated allergy to a cerebral aneurysm clip: case report.
Neurosurgery. 1998 Nov;43(5):1209-11. doi: 10.1097/00006123-199811000-00110.
10
Electrochemical and surface characterization of a nickel-titanium alloy.
Biomaterials. 1998 Apr-May;19(7-9):761-9. doi: 10.1016/s0142-9612(97)00210-x.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验