A flag-type atom shutter based on a rotating lever that is driven by a bender piezoelectric actuator was developed to manipulate atomic beams. The shutter flag was displaced by ∼10 mm to open and close a 5-mm-diameter aperture with a shutter time of 13 ms that produced small mechanical vibrations. The short-term shutter time stability for each cycle was 0.03 ms and the long-term stability over an average of 20 000 cycles was 0.02 ms. The operational cycle number (lifetime) of the shutter reached 2.0 × 10 cycles after an intermittent operation over a period of eight months in an ultra-high vacuum chamber, and another shutter in an atmospheric environment swung for 2.6 × 10 cycles of continuous operation at 5 Hz for a period of 60 days without major problems. The shutter was shown to be compatible with the operation in an ultra-high vacuum at a low 10 Pa level after a gentle baking treatment.
为了操控原子束,开发了一种基于由弯曲压电致动器驱动的旋转杠杆的旗型原子快门。快门旗移动约10毫米以打开和关闭一个直径5毫米的孔径,快门时间为13毫秒,产生的机械振动较小。每个周期的短期快门时间稳定性为0.03毫秒,平均20000个周期的长期稳定性为0.02毫秒。在超高真空腔中经过八个月的间歇运行后,快门的运行周期数(寿命)达到2.0×10次循环,另一个在大气环境中的快门以5赫兹连续运行2.6×10次循环,持续60天,没有出现重大问题。经过温和的烘烤处理后,该快门被证明与10帕斯卡低真空度的超高真空环境中的操作兼容。