Accuracy in Blood Pressure Monitoring: The Effect of Noninvasive Blood Pressure Cuff Inflation on Intra-arterial Blood Pressure Values.

作者信息

Sheshadri Veena, Tiwari Akhilesh Kumar, Nagappa Mahesh, Venkatraghavan Lashmi

机构信息

Department of Anesthesia, Toronto Western Hospital, Toronto, Canada.

Department of Anesthesia and Perioperative Medicine, University of Western Ontario, Ontario, Canada.

出版信息

Anesth Essays Res. 2017 Jan-Mar;11(1):169-173. doi: 10.4103/0259-1162.181430.

Abstract

CONTEXT

Both invasive and noninvasive blood pressure (invasive arterial blood pressure [IABP] and noninvasive BP [NIBP]) monitors are used perioperatively; however, they often produce different values. The reason for this discrepancy is not clear, and it is possible that the act of cuff inflation itself might affect the IABP values, especially with the recurrent cycling of NIBP cuff.

AIM

The aim of this study was to determine the effect of ipsilateral NIBP cuff inflation on the contralateral IABP values.

SETTINGS AND DESIGNS

Prospective, observational study.

MATERIALS AND METHODS

One hundred consecutive patients were studied. The NIBP device was set to cycle every 5 min for a total of 6 times. During each cuff inflation cycle, changes in IABP values from the arterial line in the contralateral arm were recorded. A total of 582 measurements were included for data analysis.

STATISTICAL ANALYSIS

Chi-square, paired -test, analysis of variance.

RESULTS

Mean (± standard deviation) changes in systolic BP (SBP), diastolic BP, and mean BP with cuff inflation were 6.7 ± 5.9, 2.6 ± 4.0, and 4.0 ± 3.9 mmHg, respectively. We observed an increase of 0-10 mmHg in SBP in majority (73.4%) of cuff inflations. The changes in IABP did not differ between the patients with or without hypertension or with the baseline SBP.

CONCLUSIONS

This study showed that there is a transient reactive rise in IABP values with NIBP cuff inflation. This is important information in the perioperative and intensive care settings, where both these measurement techniques are routinely used. The exact mechanism for this effect is not known but may be attributed to the pain and discomfort from cuff inflation.

摘要

背景

有创和无创血压监测(有创动脉血压[IABP]和无创血压[NIBP])在围手术期均有使用;然而,它们常常得出不同的值。这种差异的原因尚不清楚,并且袖带充气本身的行为可能会影响IABP值,尤其是在NIBP袖带反复循环充气时。

目的

本研究的目的是确定同侧NIBP袖带充气对 contralateral IABP值的影响。

设置与设计

前瞻性观察性研究。

材料与方法

对连续100例患者进行研究。NIBP设备设置为每5分钟循环一次,共循环6次。在每个袖带充气循环期间,记录对侧手臂动脉导管处IABP值的变化。总共纳入582次测量用于数据分析。

统计分析

卡方检验、配对t检验、方差分析。

结果

袖带充气时收缩压(SBP)、舒张压和平均血压的平均(±标准差)变化分别为6.7±5.9、2.6±4.0和4.0±3.9 mmHg。我们观察到在大多数(73.4%)袖带充气时SBP升高0 - 10 mmHg。有高血压或无高血压患者以及不同基线SBP患者的IABP变化无差异。

结论

本研究表明,NIBP袖带充气时IABP值会出现短暂的反应性升高。这在围手术期和重症监护环境中是重要信息,因为这两种测量技术都经常使用。这种效应的确切机制尚不清楚,但可能归因于袖带充气引起的疼痛和不适。