Aparicio Irene, Martín Julia, Santos Juan Luis, Malvar José Luis, Alonso Esteban
Department of Analytical Chemistry, Escuela Politécnica Superior, University of Seville, C/Virgen de África 7, E-41011 Seville, Spain.
Department of Analytical Chemistry, Escuela Politécnica Superior, University of Seville, C/Virgen de África 7, E-41011 Seville, Spain.
J Chromatogr A. 2017 Jun 2;1500:43-52. doi: 10.1016/j.chroma.2017.04.007. Epub 2017 Apr 7.
An analytical method based on stir bar sorptive extraction (SBSE) was developed and validated for the determination of environmental concern pollutants in environmental waters by liquid chromatography-tandem mass spectrometry (LC-MS/MS). Target compounds include six water and oil repellents (perfluorinated compounds), four preservatives (butylated hydroxytoluene and three parabens), two plasticizers (bisphenol A and di(2-ethylhexyl)phthalate), seven surfactants (four linear alkylbenzene sulfonates, nonylphenol and two nonylphenol ethoxylates), a flame retardant (hexabromocyclododecane), four hormones, fourteen pharmaceutical compounds, an UV-filter (2-ethylhexyl 4-methoxycinnamate) and nine pesticides. To achieve the simultaneous extraction of polar and non-polar pollutants two stir bar coatings were tested, the classic polydimethylsiloxane (PDMS) coating and the novel ethylene glycol modified silicone (EG-silicone). The best extraction recoveries were obtained using EG-silicone coating. The effects of sample pH, volume and ionic strength and extraction time on extraction recoveries were evaluated. The analytical method was validated for surface water and tap water samples. The method quantification limits ranged from 7.0ngL to 177ngL. The inter-day precision, expressed as relative standard deviation, was lower than 20%. Accuracy, expressed as relative recovery values, was in the range from 61 to 130%. The method was applied for the determination of the 48 target compounds in surface and tap water samples.
建立了一种基于搅拌棒吸附萃取(SBSE)的分析方法,并通过液相色谱-串联质谱(LC-MS/MS)对环境水体中的环境关注污染物进行测定,该方法经过验证。目标化合物包括六种拒水拒油剂(全氟化合物)、四种防腐剂(丁基化羟基甲苯和三种对羟基苯甲酸酯)、两种增塑剂(双酚A和邻苯二甲酸二(2-乙基己基)酯)、七种表面活性剂(四种直链烷基苯磺酸盐、壬基酚和两种壬基酚乙氧基化物)、一种阻燃剂(六溴环十二烷)、四种激素、十四种药物化合物、一种紫外线过滤剂(2-乙基己基4-甲氧基肉桂酸酯)和九种农药。为了同时萃取极性和非极性污染物,测试了两种搅拌棒涂层,即经典的聚二甲基硅氧烷(PDMS)涂层和新型的乙二醇改性硅氧烷(EG-硅氧烷)。使用EG-硅氧烷涂层获得了最佳萃取回收率。评估了样品pH值、体积、离子强度和萃取时间对萃取回收率的影响。该分析方法在地表水和自来水样品中得到验证。方法定量限为7.0ng/L至177ng/L。以相对标准偏差表示的日间精密度低于20%。以相对回收率值表示的准确度在61%至130%范围内。该方法用于地表水和自来水样品中48种目标化合物的测定。