Corneal Dose Reduction Using a Bismuth-Coated Latex Shield over the Eyes During Brain SPECT/CT.

作者信息

Matsutomo Norikazu, Fukunaga Masaaki, Onishi Hideo, Yamamoto Tomoaki

机构信息

Department of Medical Radiological Technology, Faculty of Health Sciences, Kyorin University, Tokyo, Japan

Department of Radiological Technology, Kurashiki Central Hospital, Okayama, Japan; and.

出版信息

J Nucl Med Technol. 2017 Sep;45(3):214-218. doi: 10.2967/jnmt.117.192849. Epub 2017 May 4.

Abstract

This study aimed to determine whether a bismuth-coated latex shield (B-shield) could protect the eyes during brain SPECT/CT. A shield containing the heavy metal bismuth (equivalent to a 0.15-mm-thick lead shield) was placed over a cylindric phantom and the eyes of a 3-dimensional brain phantom filled with Tc solution. Subsequently, phantoms with and without the B-shield were compared using SPECT/CT. The CT parameters were 30-200 mA and 130 kV. The dose reduction achieved by the B-shield was measured using a pencil-shaped ionization chamber. The protective effects of the B-shield were determined by evaluating relative radioactivity concentration as well as artifacts (changes in CT number), linear attenuation coefficients, and coefficients of variation on SPECT images. The radiation doses with and without the B-shield were 0.14-0.77 and 0.36-1.93 mGy, respectively, and the B-shield decreased the average radiation dose by about 60%. The B-shield also increased the mean CT number, but only at locations just beneath the surface of the phantom. Streaks of higher density near the underside of the B-shield indicated beam hardening. Linear attenuation coefficients and the coefficients of variation did not significantly differ between phantoms with and without the B-shield, and the relative Tc radioactivity concentrations were not affected. The B-shield decreased the radiation dose without affecting estimated attenuation correction or radioactivity concentrations. Although surface artifacts increased with the B-shield, the quality of the SPECT images was acceptable. B-shields can help protect pediatric patients and patients with eye diseases who undergo SPECT imaging.

摘要

本研究旨在确定铋涂层乳胶防护屏(B防护屏)在脑部单光子发射计算机断层扫描/计算机断层扫描(SPECT/CT)过程中能否保护眼睛。将一块含有重金属铋(相当于0.15毫米厚的铅防护屏)的防护屏放置在圆柱形体模以及一个充满锝溶液的三维脑部体模的眼睛上方。随后,使用SPECT/CT对有和没有B防护屏的体模进行比较。CT参数为30 - 200毫安和130千伏。使用铅笔形电离室测量B防护屏实现的剂量降低。通过评估相对放射性浓度以及伪影(CT值变化)、线性衰减系数和SPECT图像上的变异系数来确定B防护屏的保护效果。有和没有B防护屏时的辐射剂量分别为0.14 - 0.77毫戈瑞和0.36 - 1.93毫戈瑞,B防护屏使平均辐射剂量降低了约60%。B防护屏还增加了平均CT值,但仅在体模表面正下方的位置。B防护屏下方附近较高密度的条纹表明存在束硬化现象。有和没有B防护屏的体模之间线性衰减系数和变异系数没有显著差异,并且相对锝放射性浓度未受影响。B防护屏降低了辐射剂量,同时不影响估计的衰减校正或放射性浓度。尽管使用B防护屏时表面伪影增加,但SPECT图像质量仍可接受。B防护屏有助于保护接受SPECT成像的儿科患者和患有眼部疾病的患者。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验