Different tip designs in modern cataract surgery have not been studied between the reported systems.
Aim of this study is to assess the efficacy of two tip designs, Intrepid® balanced tip (BT) and Kelman tapered tip (TT), in femtosecond laser assisted cataract surgery (FLACS) and the microcoaxial torsional phacoemulsification.
Prospective randomized unmasked cohort outcome study (hospital setting).
Threehundert-forty-three eyes of 343 patientes underwent cataract surgery.
Data of n = 196 FLACS and n = 147 manual phacoemulsification were analysed. Intrepid® balanced tip and Kelman tapered tip, Alcon, USA, were tested in FLACS (LenSx Alcon, USA) und manual phaco (Alcon Centurion System, USA). Four study cohorts were formed; FLACS BT (n = 90, 70.2 years), FLACS TT (n = 106, 68.1years), Manual BT (n = 70, 71.3 years), Manual TT (n = 77, 71.8 years). The nuclei were graded in Lens Opacities Classification System (LOCS) II,III and IV.
Cumulative dissipated energy (CDE%s), balanced salt solution volume (BSSml), total longitudinal energy (%s) and torsional amplitude (%s).
Lower CDE values were seen in FLACS compared to manual phacoemulsification (CDE %s median FLACS BT 3.28, FLACS TT 4.07, Manual BT 5.57, Manual TT 6.27). There was a significant difference between CDE FLACS BT and FLACS TT (p = 0.038), and between FLACS BT and Manual TT (p = 0.001).
The right choice of tip designs in advanced phacoemulsification systems is a key factor in increasing efficacy in cataract surgery. The balanced tip showed a considerable energy-saving advantage in FLACS and manual phacoemulsification compared to the tapered tip.
现代白内障手术中的不同尖端设计在报道的系统中尚未得到研究。
本研究的目的是评估两种尖端设计(Intrepid®平衡尖端(BT)和 Kelman 渐细尖端(TT))在飞秒激光辅助白内障手术(FLACS)和微同轴扭转超声乳化术中的效果。
前瞻性随机非盲队列结局研究(医院环境)。
343 名患者的 343 只眼接受了白内障手术。
分析了 n=196 例 FLACS 和 n=147 例手动超声乳化的数据。Intrepid®平衡尖端和 Kelman 渐细尖端(美国爱尔康)分别在 FLACS(美国爱尔康 LenSx)和手动超声乳化(美国爱尔康 Centurion 系统)中进行了测试。形成了四个研究队列;FLACS BT(n=90,70.2 岁),FLACS TT(n=106,68.1 岁),手动 BT(n=70,71.3 岁),手动 TT(n=77,71.8 岁)。核被分级为晶状体混浊分类系统(LOCS)II、III 和 IV 级。
累积耗散能量(CDE%s)、平衡盐溶液量(BSSml)、总纵向能量(%)和扭转幅度(%)。
与手动超声乳化相比,FLACS 的 CDE 值较低(FLACS BT 的 CDE %s 中位数为 3.28,FLACS TT 为 4.07,手动 BT 为 5.57,手动 TT 为 6.27)。FLACS BT 和 FLACS TT 之间(p=0.038)以及 FLACS BT 和手动 TT 之间(p=0.001)存在显著差异。
在先进的超声乳化系统中选择正确的尖端设计是提高白内障手术效果的关键因素。与渐细尖端相比,平衡尖端在 FLACS 和手动超声乳化术中具有显著的节能优势。