Suppr超能文献

Port-A-Cath移除过程中残留碎片的发生率及结局

Incidence and outcome of retained Port-A-Cath fragments during removal.

作者信息

Aworanti Olugbenga Michael, Linnane Niall, Tareen Farhan, Mortell Alan

机构信息

Children's University Hospital, Temple street, Dublin, Ireland.

Our Lady's Children's Hospital, Crumlin, Dublin, Ireland.

出版信息

Pediatr Surg Int. 2017 Jul;33(7):777-781. doi: 10.1007/s00383-017-4103-6. Epub 2017 Jun 5.

Abstract

PURPOSE

Port-A-Cath devices are frequently used for long-term venous access. We postulate that long-term use predisposes them to getting stuck and retained in a central vein at the time of removal. We aim to report the incidence and outcome of this complication.

METHODS

Between January 2006 and July 2016, a retrospective review of all Port-A-Cath removals that were performed at our centre was conducted. At the time of removal, catheters that could not be removed from the vein were considered retained.

RESULTS

During the study period, 107 children had 174 episodes of silicone Sitimplant (Vygon, Ecouen, France) Port-A-Cath insertions. These children required 135 removal episodes and there were 3 (2.2%) instances whereby the catheter fragment was retained. These episodes of retained catheters only occurred in children with factor VIII deficiency (4.1% incidence in this cohort). For each episode of catheter insertion and removal, the catheters had been left in situ for a mean duration of 43 months in children with factor deficiency and no retained fragments, and the mean duration was 91 months in children with factor deficiency and retained catheter fragments (p = 0.0011).

CONCLUSIONS

Port-A-Caths that are retained after attempted removal is a complication encountered predominantly in catheters that have been in use for a prolonged duration. Furthermore, factor replacement therapy in haemophiliacs may be a risk factor for this complication.

摘要

目的

输液港装置常用于长期静脉通路。我们推测长期使用会使其在拔除时易于卡在中央静脉并残留。我们旨在报告这种并发症的发生率和结局。

方法

对2006年1月至2016年7月在我们中心进行的所有输液港拔除操作进行回顾性研究。拔除时,无法从静脉中拔出的导管被视为残留。

结果

在研究期间,107名儿童进行了174次法国威高公司生产的硅胶材质输液港植入。这些儿童需要进行135次拔除操作,有3例(2.2%)出现导管碎片残留。这些导管残留事件仅发生在患有血友病A的儿童中(该队列中的发生率为4.1%)。对于每次导管植入和拔除操作,在无导管碎片残留的血友病儿童中,导管平均留置时间为43个月,而在有导管碎片残留的血友病儿童中,平均留置时间为91个月(p = 0.0011)。

结论

拔除尝试后残留的输液港是一种主要在长期使用的导管中遇到的并发症。此外,血友病患者的因子替代治疗可能是这种并发症的一个危险因素。