CALIPER Program, Pediatric Laboratory Medicine, The Hospital for Sick Children, Toronto, ON, Canada.
Department of Laboratory Medicine and Pathobiology, University of Toronto, Toronto, ON, Canada.
Clin Chem Lab Med. 2018 Jan 26;56(2):327-340. doi: 10.1515/cclm-2017-0349.
Accurate reference intervals (RIs) based on a healthy pediatric population are essential for pediatric test result interpretation. The CALIPER project has recruited a large healthy cohort and completed a series of a priori studies to address gaps in pediatric RIs. As immunoassays from different manufacturers for endocrine and special chemistry markers are not standardized and show marked intermethod differences, direct RI studies are needed for each major analytical platform. Here, we report age- and sex-specific pediatric RIs for 29 immunoassays on the Ortho Clinical Diagnostics (Ortho) VITROS® 5600 analyzer.
Health information and blood samples were collected from healthy pediatric subjects. Using the Ortho VITROS 5600 Integrated System MicroWell Technology, 29 biomarkers were measured. Analyte concentrations were partitioned by age and sex according to the Harris and Boyd method. After removing outliers, age- and sex-specific RIs and corresponding 90% confidence intervals were calculated according to CLSI guidelines.
All analytes required age partitioning except β-human chorionic gonadotropin (β-hCG), cancer antigen 15-3 (CA15-3), rubella immunoglobulin G (rubella IgG), and vitamin D. Several analytes including estradiol, progesterone, testosterone, follicle-stimulating hormone (FSH), luteinizing hormone (LH), free triiodothyronine (FT3), total triiodothyronine (TT3), total thyroxine (TT4), thyroid uptake, ferritin, intact parathyroid hormone (iPTH), total prostate-specific antigen (tPSA), free prostate-specific antigen (fPSA), cancer antigen 125 (CA125), creatine kinase MB (CK-MB), and myoglobin showed sex differences, observed mostly with the onset of puberty.
Complex reference value trends were observed across the pediatric age range for several biomarkers examined on Ortho VITROS immunoassays. The availability of VITROS immunoassay RIs will enable accurate laboratory test interpretation and diagnosis for the pediatric population. As recommended by the CLSI EP28-A3c guidelines, implementation of these RIs should be validated for each laboratory's local pediatric population.
基于健康儿科人群的准确参考区间(RI)对于儿科检验结果的解释至关重要。CALIPER 项目已经招募了大量健康队列,并完成了一系列事先研究,以解决儿科 RI 中的差距。由于来自不同制造商的内分泌和特殊化学标志物的免疫测定未标准化,并且显示出明显的方法间差异,因此每个主要分析平台都需要进行直接 RI 研究。在这里,我们报告了 29 种免疫测定在 Ortho 临床诊断(Ortho)VITROS®5600 分析仪上的年龄和性别特异性儿科 RI。
从健康儿科受试者收集健康信息和血液样本。使用 Ortho VITROS 5600 集成系统微孔技术,测量了 29 种生物标志物。根据 Harris 和 Boyd 方法,根据年龄和性别对分析物浓度进行分区。去除离群值后,根据 CLSI 指南计算年龄和性别特异性 RI 及其相应的 90%置信区间。
除β-人绒毛膜促性腺激素(β-hCG)、癌抗原 15-3(CA15-3)、风疹 IgG(风疹 IgG)和维生素 D 外,所有分析物均需要年龄分区。包括雌二醇、孕酮、睾酮、卵泡刺激素(FSH)、黄体生成素(LH)、游离三碘甲状腺原氨酸(FT3)、总三碘甲状腺原氨酸(TT3)、总甲状腺素(TT4)、甲状腺摄取、铁蛋白、完整甲状旁腺激素(iPTH)、总前列腺特异性抗原(tPSA)、游离前列腺特异性抗原(fPSA)、癌抗原 125(CA125)、肌酸激酶 MB(CK-MB)和肌红蛋白在内的几种分析物显示出性别差异,主要发生在青春期。
在 Ortho VITROS 免疫测定上检查的几种生物标志物在儿科年龄范围内观察到复杂的参考值趋势。VITROS 免疫测定 RI 的可用性将使儿科人群的实验室检验解释和诊断更加准确。根据 CLSI EP28-A3c 指南的建议,应针对每个实验室的本地儿科人群验证这些 RI 的实施。