Comput Assist Surg (Abingdon). 2017 Dec;22(sup1):54-60. doi: 10.1080/24699322.2017.1379136. Epub 2017 Sep 22.
The effective dose (H) and organ or tissue equivalent dose (H) of a Rando phantom undergoing two brain computed tomography (CT) examination protocols were evaluated using thermoluminescent dosimeters (TLD-100H) and dose length product (DLP) methods.
TLDs were inserted into the correlated positions of an organ or tissue of Rando phantom, such as thyroid, brain, and salivary gland, using (A) axial scan: scanning the maxillae ranging from external auditory meatus to the parietal bone, and (B) helical scan: scanning from the mandible to the parietal bone. CT examinations were performed on a Philips computer tomography (Brilliance CT) at Lukang Christian Hospital. TLDs were measured using a Harshaw 3500 TLD reader. The HT of organ and tissue during the two protocols was discussed.
H were calculated using ICRP 60 and 103 at 2.67 ± 0.18 and 1.89 ± 0.23 mSv based on an axial scan, and 4.70 ± 0.38 and 4.39 ± 0.37 mSv based on a helical scan, respectively. In the DLP method, H was estimated from CTDI that was recorded directly from the console display of the CT unit and then calculated using AAPM 96. Finally, experimental results are compared with those in literature. Radiologists should choose and adjust protocols to prevent unnecessary radiation to patients and satisfying the as low as reasonably achievable (ALARA) principle. These findings will be valuable to patients, physicians, radiologists, and the public.
使用热释光剂量计(TLD-100H)和剂量长度乘积(DLP)方法评估 Rando 体模进行两次脑部计算机断层扫描(CT)检查时的有效剂量(H)和器官或组织当量剂量(H)。
将 TLD 插入 Rando 体模的相关器官或组织位置,例如甲状腺、大脑和唾液腺,使用(A)轴向扫描:从外耳道口到顶骨扫描上颌骨,以及(B)螺旋扫描:从下颌骨到顶骨扫描。CT 检查在 Lukang Christian Hospital 的飞利浦计算机断层扫描(Brilliance CT)上进行。使用 Harshaw 3500 TLD 读取器测量 TLD。讨论了两种方案中器官和组织的 HT。
根据轴向扫描,使用 ICRP 60 和 103 计算出的 H 分别为 2.67±0.18 和 1.89±0.23 mSv,根据螺旋扫描,H 分别为 4.70±0.38 和 4.39±0.37 mSv。在 DLP 方法中,从控制台显示直接从 CT 机记录的 CTDI 中估算 H,并使用 AAPM 96 进行计算。最后,将实验结果与文献进行比较。放射科医生应选择和调整方案,以防止对患者进行不必要的辐射,并满足尽可能合理地降低(ALARA)原则。这些发现对患者、医生、放射科医生和公众都将具有重要意义。