To investigate the ultrastructural features of anterior capsulotomy performed with a thermal device, the precision pulse capsulotomy (PPC).
Prospective, multicenter case series.
Consecutive patients undergoing capsulotomy with the PPC device.
Prospective study of patients undergoing capsulotomy with the PPC by 2 surgeons, followed up by routine phacoemulsification cataract surgery, was undertaken. All capsulotomy specimens were collected for scanning electron microscopy (SEM). Observations were made regarding uniformity of the capsular edge and the presence of irregularities that may compromise integrity. Comparisons were made with manual continuous curvilinear capsulorrhexis.
Ultrastructural features of PPC and presence of irregularities.
Frayed appearance of the anterior capsule edge was noted in postoperative visits under slit-lamp examination. Scanning electron microscopy sampling showed a generally uniform rolled capsular edge, but interspersed with areas of irregularity with frayed appearance at the capsule margin.
The PPC device is capable of creating reproducible, central, and precise circular capsulotomy. The ultrastructural features in ex vivo human capsulotomy specimens generally show eversion of the capsulotomy edge, but in some cases, this was accompanied by areas of irregular capsule margin with frayed edges, likely caused by dissipated thermal energy. The postoperative appearance and SEM features warrant further assessment of the PPC integrity and clinical correlation.
研究使用热设备进行前囊切开术的超微结构特征,即精密脉冲囊切开术(PPC)。
前瞻性、多中心病例系列研究。
连续接受 PPC 设备囊切开术的患者。
由 2 名外科医生对接受 PPC 囊切开术的患者进行前瞻性研究,随后进行常规的超声乳化白内障手术。所有囊切开标本均用于扫描电子显微镜(SEM)检查。观察囊边缘的均匀性和可能影响完整性的不规则性。与手动连续曲线囊切开术进行比较。
PPC 的超微结构特征和不规则性的存在。
在裂隙灯检查的术后随访中,观察到前囊边缘呈磨损状。扫描电子显微镜取样显示,一般情况下,囊边缘呈均匀的卷边,但也存在一些不规则区域,囊边缘呈磨损状。
PPC 设备能够实现可重复、中央和精确的圆形囊切开术。在体外人囊切开标本的超微结构特征中,通常显示囊切开边缘外翻,但在某些情况下,这伴随着囊边缘有不规则的磨损区域,可能是由于热能消散所致。术后外观和 SEM 特征需要进一步评估 PPC 的完整性和临床相关性。