通过连续热稀释法测量人体门静脉血流量。

Measurement of human portal blood flow by continuous thermodilution.

作者信息

Kinoshita H, Sakai K, Kubo S, Yamazaki O

机构信息

Department of Surgery, Osaka City University Medical School, Japan.

出版信息

J Surg Res. 1989 Mar;46(3):235-40. doi: 10.1016/0022-4804(89)90063-2.

Abstract

We wanted to use the Ganz thermistor catheter, developed to measure blood flow in human coronary sinuses, to measure portal venous flow, which is greater. We prepared a model circuit with a high flow rate and studied the method for suitability and reproducibility of the measurements. Correlation was close between the measured and calculated flow rates when the thermal indicator was injected at the rate of 48 ml/min. Measurements were highly reproducible. The catheter was then tested clinically. It was inserted into the portal trunk during percutaneous transhepatic catheterization to measure the portal venous flow. Measurements in patients with cirrhosis of the liver were close to those reported by other workers, and were reproducible. The results suggest that this method will be useful for the continuous local measurement of portal venous flow that varies with time, and will provide information about portal circulation in hepatic disease.

摘要

我们想用为测量人体冠状窦血流而研发的甘兹热敏电阻导管来测量更大的门静脉血流。我们制备了一个高流速的模型电路,并研究了测量的适用性和可重复性方法。当以48毫升/分钟的速率注入热指示剂时,测量流速与计算流速之间的相关性密切。测量具有高度可重复性。然后对该导管进行了临床测试。在经皮肝穿刺导管插入术中,将其插入门静脉主干以测量门静脉血流。肝硬化患者的测量结果与其他研究人员报告的结果相近,且具有可重复性。结果表明,该方法将有助于对随时间变化的门静脉血流进行连续局部测量,并将提供有关肝病门静脉循环的信息。