The Reliability of Standing Sagittal Measurements of Spinal Curvature and Range of Motion in Older Women With and Without Hyperkyphosis Using a Skin-Surface Device.

作者信息

Roghani Tayebeh, Khalkhali Zavieh Minoo, Rahimi Abbas, Talebian Saeed, Dehghan Manshadi Farideh, Akbarzadeh Baghban Alireza, King Nicole, Katzman Wendy

机构信息

Department of Physiotherapy, School of Rehabilitation, Shahid Beheshti University of Medical Sciences, Tehran, Iran.

Department of Physiotherapy, School of Rehabilitation, Shahid Beheshti University of Medical Sciences, Tehran, Iran.

出版信息

J Manipulative Physiol Ther. 2017 Nov-Dec;40(9):685-691. doi: 10.1016/j.jmpt.2017.07.008.

Abstract

OBJECTIVE

The purpose of this study was to investigate the intrarater reliability of a skin-surface instrument (Spinal Mouse, Idiag, Voletswil, Switzerland) in measuring standing sagittal curvature and global mobility of the spine in older women with and without hyperkyphosis.

METHODS

Measurements were made in 19 women with hyperkyphosis (thoracic kyphosis angle ≥50°), mean age 67 ± 5 years, and 14 women without hyperkyphosis (thoracic kyphosis angle <50°), mean age 63 ± 6 years. Sagittal thoracic and lumbar curvature and mobility of the spine were assessed with the Spinal Mouse during neutral standing, full spinal flexion, and full spinal extension. Tests were performed by the same examiner on 2 days with a 72-hour interval. The intrarater reliability of the measurements was analyzed using the intraclass correlation coefficient, standard error of measurement and minimal detectable change.

RESULTS

Intraclass correlation coefficients ranged from 0.89 to 0.99 in both groups. The standard errors of measurement ranged from 1.02° to 2.06° in the hyperkyphosis group and from 1.15° to 2.22° in the normal group. The minimal detectable change ranged from 2.85° to 5.73° in the hyperkyphosis group and from 3.20° to 6.17° in the normal group.

CONCLUSIONS

Our results indicated that the Spinal Mouse has excellent intrarater reliability for the measurement of sagittal thoracic and lumbar curvature and mobility of the spine in older women.

摘要

目的

本研究旨在调查一种皮肤表面仪器(Spinal Mouse,Idiag,瑞士沃莱茨维尔)在测量患有和未患有脊柱后凸的老年女性脊柱矢状面弯曲度和整体活动度时的评分者内信度。

方法

对19名患有脊柱后凸(胸椎后凸角≥50°)、平均年龄67±5岁的女性和14名未患有脊柱后凸(胸椎后凸角<50°)、平均年龄63±6岁的女性进行测量。在中立位站立、脊柱完全前屈和脊柱完全后伸时,使用Spinal Mouse评估胸椎和腰椎矢状面弯曲度以及脊柱活动度。测试由同一名检查者在间隔72小时的两天内进行。使用组内相关系数、测量标准误差和最小可检测变化分析测量的评分者内信度。

结果

两组的组内相关系数范围均为0.89至0.99。脊柱后凸组的测量标准误差范围为1.02°至2.06°,正常组为1.15°至2.22°。脊柱后凸组的最小可检测变化范围为2.85°至5.73°,正常组为3.20°至6.17°。

结论

我们的结果表明,Spinal Mouse在测量老年女性胸椎和腰椎矢状面弯曲度以及脊柱活动度方面具有出色的评分者内信度。