Halstuch Daniel, Baniel Jack, Lifshitz David, Sela Sivan, Ber Yaara, Margel David
1 Division of Urology, Rabin Medical Center , Petah-Tikva, Israel .
2 Division of Surgery, Sackler Faculty of Medicine, Tel Aviv University , Tel Aviv, Israel .
J Endourol. 2018 Mar;32(3):252-256. doi: 10.1089/end.2017.0694. Epub 2018 Jan 26.
To measure needle tip deflection during transrectal ultrasound (TRUS) prostate biopsy and evaluate predictors for needle tip deflection.
Analysis of 568 prostate biopsies obtained from 51 consecutive patients who underwent a standard 12-core TRUS guided prostate biopsy. TRUS guided prostate biopsies were performed using BK flex500, with a side-fire biplane probe. Each biopsy core image was captured and clinical data were recorded prospectively. The angle between the expected trajectory of the needle and actual needle course was measured using the longitudinal view of the captured image. The distance between expected and actual needle tip was calculated. We measured median and interquartile needle tip deflection rate stratified by side and location (apex, midgland, base). Univariable and multivariable linear regressions analysis were performed.
The overall median needle tip deflection was 1.77 mm (IQR 1.35-2.47). Location did not significantly alter needle deflection measurements. On multivariable linear regression analysis, higher prostate volume (B = 0.007 95%, CI 0.004, 0.011; p < 0.001) and the right sided biopsy (B = 0.191 95%, CI 0.047, 0.336; p = 0.010) emerged as predictors of higher needle tip deflection.
To the best of our knowledge this is the first study to measure needle tip deflection during TRUS guided prostate biopsies. We demonstrated that larger prostate size and biopsy side may affect the accuracy of biopsies. These results may have clinical implication to those performing targeted biopsies.
测量经直肠超声(TRUS)引导下前列腺穿刺活检时的针尖偏移,并评估针尖偏移的预测因素。
对51例连续接受标准12针TRUS引导下前列腺穿刺活检患者的568次前列腺穿刺活检进行分析。使用BK flex500侧射双平面探头进行TRUS引导下前列腺穿刺活检。前瞻性地采集每次穿刺活检核心图像并记录临床数据。利用采集图像的纵视图测量针的预期轨迹与实际针道之间的角度。计算预期针尖与实际针尖之间的距离。我们按穿刺部位(尖部、中部、基部)和穿刺侧别分层测量针尖偏移率的中位数和四分位数间距。进行单变量和多变量线性回归分析。
总体针尖偏移中位数为1.77mm(四分位数间距1.35 - 2.47)。穿刺部位对针尖偏移测量无显著影响。多变量线性回归分析显示,前列腺体积较大(B = 0.007,95%置信区间0.004,0.011;p < 0.001)和右侧穿刺活检(B = 0.191,95%置信区间0.047,0.336;p = 0.010)是针尖偏移较大的预测因素。
据我们所知,这是第一项测量TRUS引导下前列腺穿刺活检时针尖偏移的研究。我们证明前列腺体积较大和穿刺侧别可能影响穿刺活检的准确性。这些结果可能对进行靶向穿刺活检的人员具有临床意义。