Suppr超能文献

使用患者相对校准方法对肝动脉栓塞术 Tc-MAA SPECT 3D 剂量学中衰减和散射校正缺失的影响:基于临床图像的回顾性研究。

Impact of missing attenuation and scatter corrections on Tc-MAA SPECT 3D dosimetry for liver radioembolization using the patient relative calibration methodology: A retrospective investigation on clinical images.

机构信息

Medical Physics Service, Department of Medical Images and Radiation Sciences, Istituto Europeo di Oncologia, 20141, Milan, Italy.

Nuclear Medicine Department, Foundation IRCCS Istituto Nazionale Tumori, 20133, Milan, Italy.

出版信息

Med Phys. 2018 Apr;45(4):1684-1698. doi: 10.1002/mp.12774. Epub 2018 Feb 23.

Abstract

PURPOSE

To investigate the clinical implication of performing pre-treatment dosimetry for Y-microspheres liver radioembolization on Tc-MAA SPECT images reconstructed without attenuation or scatter correction and quantified with the patient relative calibration methodology.

METHODS

Twenty-five patients treated with SIR-Spheres at Istituto Europeo di Oncologia and 31 patients treated with TheraSphere at Istituto Nazionale Tumori were considered. For each acquired Tc-MAA SPECT, four reconstructions were performed: with attenuation and scatter correction (AC_SC), only attenuation (AC_NoSC), only scatter (NoAC_SC) and without corrections (NoAC_NoSC). Absorbed dose maps were calculated from the activity maps, quantified applying the patient relative calibration to the SPECT images. Whole Liver (WL) and Tumor (T) regions were drawn on CT images. Injected Liver (IL) region was defined including the voxels receiving absorbed dose >3.8 Gy/GBq. Whole Healthy Liver (WHL) and Healthy Injected Liver (HIL) regions were obtained as WHL = WL - T and HIL = IL - T. Average absorbed dose to WHL and HIL were calculated, and the injection activity was derived following each Institute's procedure. The values obtained from AC_NoSC, NoAC_SC and NoAC_NoSC images were compared to the reference value suggested by AC_SC images using Bland-Altman analysis and Wilcoxon paired test (5% significance threshold). Absorbed-dose maps were compared to the reference map (AC_SC) in global terms using the Voxel Normalized Mean Square Error (%VNMSE), and at voxel level by calculating for each voxel the normalized difference with the reference value. The uncertainty affecting absorbed dose at voxel level was accounted for in the comparison; to this purpose, the voxel counts fluctuation due to Poisson and reconstruction noise was estimated from SPECT images of a water phantom acquired and reconstructed as patient images.

RESULTS

NoAC_SC images lead to activity prescriptions not significantly different from the reference AC_SC images; the individual differences (<0.1 GBq for all IEO patients, <0.6 GBq for all but one INT patients) were comparable to the uncertainty affecting activity measurement. AC_NoSC and NoAC_NoSC images, instead, yielded significantly different activity prescriptions and wider 95% confidence intervals in the Bland-Altman analysis. Concerning the absorbed dose map, AC_NoSC images had the smallest %VNMSE value and the highest fraction of voxels differing less than 2 standard deviations from AC_SC.

CONCLUSIONS

The patient relative calibration methodology can compensate for the missing attenuation correction when performing healthy liver pre-treatment dosimetry: safe treatments can be planned even on NoAC_SC images, suggesting activities comparable to AC_SC images. Scatter correction is recommended due to its heavy impact on healthy liver dosimetry.

摘要

目的

研究在未进行衰减或散射校正的 Tc-MAA SPECT 图像上重建时,使用患者相对校准方法对 Y-微球肝脏放射性栓塞进行预处理剂量测定的临床意义。

方法

考虑了在欧洲肿瘤研究所接受 SIR-Spheres 治疗的 25 名患者和在国家肿瘤研究所接受 TheraSphere 治疗的 31 名患者。对于每个采集的 Tc-MAA SPECT,进行了四种重建:带有衰减和散射校正 (AC_SC)、仅衰减 (AC_NoSC)、仅散射 (NoAC_SC) 和无校正 (NoAC_NoSC)。从活性图中计算吸收剂量图,并通过将患者相对校准应用于 SPECT 图像来对其进行量化。在 CT 图像上绘制全肝 (WL) 和肿瘤 (T) 区域。注入肝 (IL) 区域的定义包括接收吸收剂量> 3.8 Gy/GBq 的体素。全健康肝 (WHL) 和健康注入肝 (HIL) 区域作为 WHL = WL - T 和 HIL = IL - T 获得。计算 WHL 和 HIL 的平均吸收剂量,并根据每个研究所的程序推导注射活性。使用 Bland-Altman 分析和 Wilcoxon 配对检验 (5%显著性阈值) 将来自 AC_NoSC、NoAC_SC 和 NoAC_NoSC 图像的值与 AC_SC 图像的参考值进行比较。使用体素归一化均方误差 (%VNMSE) 从全局角度比较吸收剂量图,并通过计算每个体素与参考值的归一化差值在体素水平上进行比较。在比较中考虑了影响体素水平吸收剂量的不确定性;为此,从患者图像采集和重建的水模体 SPECT 图像中估计了由于泊松和重建噪声导致的体素计数波动。

结果

NoAC_SC 图像导致的活性处方与参考 AC_SC 图像没有显著差异;个体差异(所有 IEO 患者<0.1GBq,所有患者<0.6GBq,但有一位患者除外)与影响活性测量的不确定性相当。AC_NoSC 和 NoAC_NoSC 图像,另一方面,在 Bland-Altman 分析中产生了显著不同的活性处方和更宽的 95%置信区间。关于吸收剂量图,AC_NoSC 图像的 %VNMSE 值最小,与 AC_SC 差异小于 2 个标准差的体素比例最高。

结论

当进行健康肝脏预处理剂量测定时,患者相对校准方法可以补偿缺失的衰减校正:即使在 NoAC_SC 图像上也可以安全地计划治疗,提示与 AC_SC 图像相比,活性相当。由于散射校正对健康肝脏剂量学有重大影响,因此建议进行散射校正。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验