State Key Laboratory of Chemical Resource Engineering, Beijing University of Chemical Technology, Beijing, China; Department of Chemistry, Beijing Key Laboratory of Microanalytical Methods and Instrumentation, MOE Key Laboratory of Bioorganic Phosphorus Chemistry & Chemical Biology, Tsinghua University, Beijing 100084, China.
Department of Chemistry, Beijing Key Laboratory of Microanalytical Methods and Instrumentation, MOE Key Laboratory of Bioorganic Phosphorus Chemistry & Chemical Biology, Tsinghua University, Beijing 100084, China.
Talanta. 2018 May 1;181:217-223. doi: 10.1016/j.talanta.2018.01.016. Epub 2018 Jan 8.
Matrix-assisted laser desorption/ionization mass spectrometry (MALDI-MS) is a fast analysis tool for the detection of a wide range of analytes. However, heterogeneous distribution of matrix/analyte cocrystal, variation in signal intensity and poor experimental reproducibility at different locations of the same spot means difficulty in quantitative analysis. In this work, carbon nanotubes (CNTs) were employed as adsorbent for analyte cum matrix on a conductive porous membrane as a novel mass target plate. The sample pretreatment step was achieved by enrichment and dead-end filtration and dried by a solid-liquid separation. This approach enables the homogeneous distribution of analyte in the matrix, good shot-to-shot reproducibility in signals and quantitative detection of peptide and protein at different concentrations with correlation coefficient (R) of 0.9920 and 0.9909, respectively. The simple preparation of sample in a short time, uniform distribution of analyte, easy quantitative detection, and high reproducibility makes this technique useful and may diversify the application of MALDI-MS for quantitative detection of a variety of proteins.
基质辅助激光解吸/电离质谱(MALDI-MS)是一种快速分析工具,可用于检测广泛的分析物。然而,基质/分析物共晶体的不均匀分布、信号强度的变化以及同一斑点不同位置的实验重现性差,意味着定量分析存在困难。在这项工作中,碳纳米管(CNTs)被用作吸附剂,将分析物与基质共晶吸附在导电多孔膜上,作为一种新型的质量靶板。通过富集和死端过滤来实现样品预处理步骤,然后通过固-液分离干燥。这种方法可实现分析物在基质中的均匀分布,信号具有良好的重现性,可对不同浓度的肽和蛋白质进行定量检测,其相关系数(R)分别为 0.9920 和 0.9909。该技术样品制备简单,耗时短,分析物分布均匀,易于定量检测,重现性高,具有很好的应用前景,可能会使 MALDI-MS 技术在各种蛋白质的定量检测方面得到更广泛的应用。