Nandy S, Tarbell J M
J Biomech Eng. 1986 Aug;108(3):228-31. doi: 10.1115/1.3138607.
The accuracy of a flush-mounted hot film anemometer probe for wall shear stress measurements in physiological pulsatile flows was evaluated in fully developed pulsatile flow in a rigid straight tube. Measured wall shear stress waveform based on steady flow anemometer probe calibrations were compared to theoretical wall shear stress waveforms based on well-established theory and measured flow rate waveforms. The measured and theoretical waveforms were in close agreement during systole (average deviation of 14 percent at peak systole). As expected, agreement was poor during diastole because of flow reversal and diminished frequency response at low shear rate.
在刚性直管中充分发展的脉动流中,评估了用于生理脉动流壁面剪应力测量的嵌入式热膜风速仪探头的准确性。将基于稳态风速仪探头校准测得的壁面剪应力波形与基于成熟理论和测得的流量波形的理论壁面剪应力波形进行了比较。在收缩期,测得的波形与理论波形吻合良好(收缩期峰值时平均偏差为14%)。正如预期的那样,在舒张期由于血流逆转和低剪应力率下频率响应降低,两者吻合较差。