Carpentier-Edwards PERIMOUNT Magna Ease 假体的流体动力学性能。

The Fluid Dynamical Performance of the Carpentier-Edwards PERIMOUNT Magna Ease Prosthesis.

机构信息

Department of Thoracic and Cardiovascular Surgery, West-German Heart and Vascular Center Essen, University Hospital Essen, Essen, Germany.

Chair of Mechanics and Robotics, University Duisburg-Essen, Campus Duisburg, Lotharstraße 1, 47057 Duisburg, Germany.

出版信息

Biomed Res Int. 2018 Jan 10;2018:5429594. doi: 10.1155/2018/5429594. eCollection 2018.

Abstract

The aim of the present in vitro study was the evaluation of the fluid dynamical performance of the Carpentier-Edwards PERIMOUNT Magna Ease depending on the prosthetic size (21, 23, and 25 mm) and the cardiac output (3.6-6.4 L/min). A self-constructed flow channel in combination with particle image velocimetry (PIV) enabled precise results with high reproducibility, focus on maximal and local peek velocities, strain, and velocity gradients. These flow parameters allow insights into the generation of forces that act on blood cells and the aortic wall. The results showed that the 21 and 23 mm valves have a quite similar performance. Maximal velocities were 3.03 ± 0.1 and 2.87 ± 0.13 m/s; maximal strain , 913.81 ± 173.25 and 896.15 ± 88.16 1/s; maximal velocity gradient , 1203.14 ± 221.84 1/s and 1200.81 ± 61.83 1/s. The 25 mm size revealed significantly lower values: maximal velocity, 2.47 ± 0.15 m/s; maximal strain , 592.98 ± 155.80 1/s; maximal velocity gradient , 823.71 ± 38.64 1/s. In summary, the 25 mm Magna Ease was able to create a wider, more homogenous flow with lower peak velocities especially for higher flow rates. Despite the wider flow, the velocity values close to the aortic walls did not exceed the level of the smaller valves.

摘要

本体外研究的目的是评估 Carpentier-Edwards PERIMOUNT Magna Ease 瓣膜的流体动力学性能,该瓣膜取决于假体尺寸(21、23 和 25mm)和心输出量(3.6-6.4L/min)。结合粒子图像测速法(PIV)的自构建流道能够实现精确且具有高重复性的结果,重点是最大和局部峰值速度、应变和速度梯度。这些流动参数可以深入了解作用于血细胞和主动脉壁的力的产生。结果表明,21 和 23mm 瓣膜的性能非常相似。最大速度分别为 3.03±0.1m/s 和 2.87±0.13m/s;最大应变为 913.81±173.25 1/s 和 896.15±88.16 1/s;最大速度梯度分别为 1203.14±221.84 1/s 和 1200.81±61.83 1/s。25mm 尺寸显示出明显较低的值:最大速度为 2.47±0.15m/s;最大应变为 592.98±155.80 1/s;最大速度梯度为 823.71±38.64 1/s。总之,25mm Magna Ease 能够产生更宽、更均匀的流动,特别是在较高流量下,峰值速度较低。尽管流量较宽,但靠近主动脉壁的速度值并未超过较小瓣膜的水平。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验