Do relevant shear forces appear in isokinetic shoulder testing to be implemented in biomechanical models?

作者信息

Bélaise Colombe, Michaud Benjamin, Boucher Mathieu, Begon Mickaël

机构信息

Laboratory of Simulation and Modelisation of Movement, Université de Montréal, Montreal, QC, Canada; Sainte-Justine Hospital Research Centre, Montreal, QC, Canada.

Laboratory of Simulation and Modelisation of Movement, Université de Montréal, Montreal, QC, Canada.

出版信息

J Biomech. 2018 Apr 27;72:241-246. doi: 10.1016/j.jbiomech.2018.02.029. Epub 2018 Mar 2.

Abstract

Isokinetic dynamometers measure joint torques about a single fixed rotational axis. Previous studies yet suggested that muscles produce both tangential and radial forces during a movement, so that the contact forces exerted to perform this movement are multidirectional. Then, isokinetic dynamometers might neglect the torque components about the two other Euclidean space axes. Our objective was to experimentally quantify the shear forces impact on the overall shoulder torque, by comparing the dynamometer torque to the torque computed from the contact forces at the hand and elbow. Ten healthy women performed isokinetic maximal internal/external concentric/eccentric shoulder rotation movements. The hand and elbow contact forces were measured using two six-axis force sensors. The main finding is that the contact forces at the hand were not purely tangential to the direction of the movement (effectiveness indexes from 0.26 ± 0.25 to 0.54 ± 0.20), such that the resulting shoulder torque computed from the two force sensors was three-dimensional. Therefore, the flexion and abduction components of the shoulder torque measured by the isokinetic dynamometer were significantly underestimated (up to 94.9%). These findings suggest that musculoskeletal models parameters should not be estimated without accounting for the torques about the three space axes.

摘要

等速测力计测量围绕单个固定旋转轴的关节扭矩。然而,先前的研究表明,肌肉在运动过程中会产生切向力和径向力,因此为执行该运动而施加的接触力是多方向的。那么,等速测力计可能会忽略另外两个欧几里得空间轴上的扭矩分量。我们的目标是通过比较测力计扭矩与根据手部和肘部接触力计算出的扭矩,实验性地量化剪切力对整体肩部扭矩的影响。十名健康女性进行了等速最大内/外同心/偏心肩部旋转运动。使用两个六轴力传感器测量手部和肘部的接触力。主要发现是,手部的接触力并非完全与运动方向相切(有效性指数从0.26±0.25到0.54±0.20),因此根据两个力传感器计算出的肩部扭矩是三维的。因此,等速测力计测量的肩部扭矩的屈曲和外展分量被显著低估(高达94.9%)。这些发现表明,在不考虑三个空间轴上的扭矩的情况下,不应估计肌肉骨骼模型参数。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验