Kennedy Joseph H, Palaty Jan, Gill Chris G, Wiseman Justin M
Prosolia Inc., Indianapolis, IN, 46202, USA.
LifeLabs Medical Laboratory Services, Burnaby, BC, V5G 4V8, Canada.
Rapid Commun Mass Spectrom. 2018 Aug 15;32(15):1280-1286. doi: 10.1002/rcm.8164.
Drug overdose deaths due to fentanyls and other novel psychoactive substances (NPS) are on the rise. The higher potencies of fentanyl analogs compared with morphine require new technologies to identify and quantitate NPS.
Paper spray tandem mass spectrometry (MS/MS) and high-resolution mass spectrometry were used to identify and measure fentanyl analogs as well as common drugs of abuse in urine samples from substance use disorder clinics. Ten-microliter urine samples were deposited directly on paper spray cartridges previously loaded with internal standards, dried, and analyzed with no other sample treatment. Quantitative results were obtained using MS/MS. Individual drugs were identified using high-resolution accurate mass spectrometry, and confirmed by data-dependent MS/MS.
Calibration curves in urine were linear over a range of 0.5-50 ng/mL with R of 0.99 or better for eight representative fentanyl analogs. Cartridges preloaded with internal standards demonstrated satisfactory quantitative results compared with LC/MS. Direct identification and confirmation of fentanyl analogs and other common drugs of abuse in urine using high-resolution accurate mass and MS/MS fragmentation were demonstrated at low picogram levels.
Paper spray mass spectrometry can reliably identify and quantitate fentanyl analogs and other drugs of abuse in urine. Using paper spray cartridges as collection devices reduces exposure and transportation risks associated with biological fluids. Cartridges preloaded with labeled internal standards can be effective for targeted screening of fentanyl analogs and other drugs of abuse.
因芬太尼和其他新型精神活性物质(NPS)导致的药物过量死亡人数正在上升。与吗啡相比,芬太尼类似物的效力更高,这就需要新技术来识别和定量NPS。
采用纸喷雾串联质谱(MS/MS)和高分辨率质谱来识别和测量来自物质使用障碍诊所的尿液样本中的芬太尼类似物以及常见的滥用药物。将10微升尿液样本直接沉积在预先装有内标的纸喷雾小柱上,干燥后无需其他样品处理即可进行分析。使用MS/MS获得定量结果。使用高分辨率精确质量质谱识别单个药物,并通过数据依赖的MS/MS进行确认。
尿液中的校准曲线在0.5 - 50 ng/mL范围内呈线性,八种代表性芬太尼类似物的R值为0.99或更高。与液相色谱/质谱法相比,预先装有内标的小柱显示出令人满意的定量结果。在低皮克水平下,使用高分辨率精确质量和MS/MS碎片技术直接鉴定和确认了尿液中的芬太尼类似物和其他常见滥用药物。
纸喷雾质谱法能够可靠地识别和定量尿液中的芬太尼类似物及其他滥用药物。使用纸喷雾小柱作为采集装置可降低与生物体液相关的暴露和运输风险。预先装有标记内标的小柱可有效用于芬太尼类似物和其他滥用药物的靶向筛查。