Yuan Di, Gao Xun, Yao Shuang, Zheng Bao-luo
Guang Pu Xue Yu Guang Pu Fen Xi. 2016 Aug;36(8):2617-20.
The elemental concentration of Cr contained in soil is measured with laser induced breakdown spectroscopy (LIBS). As the LIBS laser power, the laser wavelength is of 1 064 nm with pulse width of 8 ns and laser repetition frequency of 10 Hz in this experiment and Cr 427.4 nm is selected as the analytical line of LIBS. The experimental results show that the relative standard deviation (RSD) of the detected content of Cr is 12.1% at the delay time of 4.78 μs and the soil sample surface 1 mm behind lens focal point. The limit of detection (LOD) of LIBS is 2.01 ppm. The measured relative deviation between the measured value and the nature value is 5.15%. Because the LIBS has advantages such as lower LOD and higher detection precision, LIBS has a guiding significance to detect heavy metal in soil and to monitor environmental quality precisely and rapidly.
采用激光诱导击穿光谱法(LIBS)测量土壤中铬的元素浓度。在本实验中,LIBS激光功率下,激光波长为1064 nm,脉冲宽度为8 ns,激光重复频率为10 Hz,选择Cr 427.4 nm作为LIBS的分析谱线。实验结果表明,在延迟时间为4.78 μs且位于透镜焦点后方1 mm处的土壤样品表面,检测到的Cr含量的相对标准偏差(RSD)为12.1%。LIBS的检测限(LOD)为2.01 ppm。测量值与真值之间的相对偏差为5.15%。由于LIBS具有检测限低、检测精度高等优点,对土壤中重金属的检测以及精确快速地监测环境质量具有指导意义。