Midwest Cardiovascular Research Foundation, Davenport, Iowa.
Division of Cardiology, Columbia Presbyterian, New York, New York.
JACC Cardiovasc Interv. 2018 Oct 8;11(19):1995-2003. doi: 10.1016/j.jcin.2018.05.025.
The WISE LE (WIRION™ EPS in Lower Extremities Arteries) study was designed to assess the clinical performance of the WIRION Embolic Protection System (EPS) in subjects undergoing lower extremity atherectomy for the treatment of peripheral artery disease.
Embolization is ubiquitous during endovascular procedures for lower extremity peripheral artery disease.
The WISE LE was a multicenter study, performed in the United States and Germany. The primary endpoint was freedom from major adverse events (MAEs) occurring within 30 days post-procedure and was compared with an objective performance goal derived from historical atherectomy trials. MAE was defined as a serious adverse event that resulted in death, acute myocardial infarction, thrombosis, pseudoaneurysm, dissection (grade C or greater), or clinical perforation at the filter location, clinically relevant distal embolism, unplanned amputation, or clinically driven target vessel revascularization. The study also included a histopathological analysis of debris captured by the filter during the procedures.
The study protocol specified enrollment of 153 patients with the primary endpoint successfully met if 18 (12.0%) or fewer MAEs occurred. A pre-specified interim analysis performed after 103 patients revealed only 2 MAEs, and the study was stopped because it had met its pre-determined metric for success. Lesion deemed not accessible by the WIRION EPS occurred in 7 patients. Debris of <1-mm, 1- to 2-mm, and >2-mm diameter were found in 98%, 22%, and 9% of patients, respectively.
The WIRION EPS is safe and noninferior to the pre-specified performance goal in capturing debris in the vast majority of patients and with the use of a broad range of atherectomy systems.
WISE LE(WIRION™ 下肢动脉中的 EPS)研究旨在评估 WIRION 栓塞保护系统(EPS)在接受下肢动脉粥样硬化切除术治疗外周动脉疾病的患者中的临床性能。
在外周动脉疾病的血管内治疗过程中,栓塞是普遍存在的。
WISE LE 是一项多中心研究,在美国和德国进行。主要终点是在术后 30 天内无重大不良事件(MAE)发生,并与从历史动脉粥样硬化切除术试验中得出的客观性能目标进行比较。MAE 定义为导致死亡、急性心肌梗死、血栓形成、假性动脉瘤、夹层(C 级或更高)或滤器位置的临床穿孔、临床相关的远端栓塞、非计划截肢或临床驱动的靶血管血运重建的严重不良事件。该研究还包括对手术过程中滤器捕获的碎片进行组织病理学分析。
研究方案规定,如果 18 例(12.0%)或更少的 MAE 发生,则主要终点成功招募 153 例患者。在 103 例患者后进行的预先指定的中期分析仅显示 2 例 MAE,研究因达到其预先确定的成功标准而停止。认为 WIRION EPS 无法到达的病变发生在 7 例患者中。<1mm、1-2mm 和>2mm 直径的碎片分别在 98%、22%和 9%的患者中发现。
WIRION EPS 在捕获绝大多数患者的碎片方面是安全的,且不劣于预先指定的性能目标,同时使用了广泛的动脉粥样硬化切除术系统。