a Joint Shantou International Eye Center , Shantou University & the Chinese University of Hong Kong , Guangdong , PR China.
b Department of Ophthalmology & Visual Sciences , The Chinese University of Hong Kong , Guangdong , PR China.
Curr Eye Res. 2019 Mar;44(3):324-328. doi: 10.1080/02713683.2018.1542735. Epub 2018 Nov 9.
Fifty-one healthy young subjects were recruited in this prospective study. MPOD parameters including maximum optical intensity (OD), mean OD, volume and area, were measured using a single 460-nm wavelength reflection method (Visucam 200, Carl Zeiss Meditec) twice both before and after pupil dilation. Paired t test, Pearson's correlation, intraclass correlation coefficient (ICC) and Bland-Altman plots were used to evaluate the agreement and correlation of MPOD parameters before and after mydriasis.
The differences of maximum OD, mean OD and volume between the two measurements without mydriasis was statistically significant for all parameters, but not significant between the two measurements with mydriasis or between non-mydriatic and mydriatic measures. The ICCs were higher for the two measurements with mydriasis (range: 0.687-0.840) than those without mydriasis (range: 0.448-0.695) or non-mydriatic and mydriatic measures (range: 0.473-0.769). The 95% limit of agreement (LOA) of measurements between the two measurements after mydriasis was the smallest among all the other comparisons. The coefficient of repeatability was lower for the measurements with mydriasis than those without mydriasis.
MPOD measurement by Visucam 200 with mydriasis has higher repeatability than the measurement without mydriasis. The non-mydriatic and mydriatic measures had less agreement compared to the two measure after mydriasis. Therefore, mydriasis is recommended.
用单波长反射光度法评估非散瞳和散瞳条件下黄斑色素光密度(MPOD)测量的重复性。
比较非散瞳测量与散瞳测量的 MPOD 值。
本前瞻性研究招募了 51 名健康年轻受试者。使用单 460nm 波长反射法(Visucam 200,Carl Zeiss Meditec)两次测量 MPOD 参数,包括最大光密度(OD)、平均 OD、体积和面积,两次测量均在瞳孔扩张前后进行。采用配对 t 检验、Pearson 相关、组内相关系数(ICC)和 Bland-Altman 图评估散瞳前后 MPOD 参数的一致性和相关性。
未经散瞳两次测量的最大 OD、平均 OD 和体积差异均有统计学意义,但散瞳后两次测量或未经散瞳和散瞳测量之间的差异无统计学意义。散瞳后两次测量的 ICC 较高(范围:0.687-0.840),而未经散瞳两次测量的 ICC 较低(范围:0.448-0.695)或未经散瞳和散瞳测量的 ICC 较低(范围:0.473-0.769)。散瞳后两次测量之间的 95%限差值(LOA)最小。散瞳后测量的重复性系数低于未经散瞳测量。
用 Visucam 200 散瞳测量的 MPOD 重复性高于非散瞳测量。与散瞳后两次测量相比,非散瞳和散瞳测量的一致性较差。因此,建议散瞳。