Kario Kazuomi, Hoshide Satoshi, Saito Kimiyo, Sato Keiko, Hamasaki Haruna, Suwa Hiromi, Tomitani Naoko
Department of Medicine, School of Medicine, Division of Cardiovascular Medicine, Jichi Medical University, Shimotsuke, Japan.
Blood Press Monit. 2019 Feb;24(1):38-41. doi: 10.1097/MBP.0000000000000357.
The aim of this study was to validate the TM-2441 ambulatory blood pressure monitoring (ABPM) device using the ISO 81060-2:2013 standard.
Participants were healthy individuals aged more than 12 years who were recruited from among the outpatients and volunteers of Jichi Medical University (Tochigi, Japan). The same-arm sequence protocol (clinical validation) and the opposite-limb simultaneous method (ambulatory validation) from the ISO 81060-2:2013 standard were used.
One hundred and seven participants were enrolled; 85 participated in the clinical validation and 35 participated in the ambulatory validation (13 participants were included in both validation protocols). The TM-2441 device performed well against the standard in both the clinical and ambulatory validations; the mean and SD values for the differences between device and observed systolic and diastolic blood pressure values in both tests fulfilled criterion 1 and criterion 2 of the standard. The Bland-Altman plots did not show any systematic variation in the error.
The TM-2441 ABPM device was accurate and fulfilled all ISO 81060-2:2013 standard requirements for ABPM determination in adults. It is therefore suitable for use for ABPM in adults with hypertension.
本研究旨在使用ISO 81060 - 2:2013标准验证TM - 2441动态血压监测(ABPM)设备。
参与者为年龄超过12岁的健康个体,从日本枥木县自治医科大学的门诊患者和志愿者中招募。采用了ISO 81060 - 2:2013标准中的同侧序列方案(临床验证)和对侧肢体同步方法(动态验证)。
共招募了107名参与者;85人参与临床验证,35人参与动态验证(13人同时参与两种验证方案)。TM - 2441设备在临床和动态验证中均符合标准;两项测试中设备与观察到的收缩压和舒张压值之间差异的均值和标准差均满足标准的标准1和标准2。Bland - Altman图未显示误差有任何系统变化。
TM - 2441 ABPM设备准确无误,满足ISO 81060 - 2:2013标准对成人ABPM测定的所有要求。因此,它适用于高血压成人患者的ABPM。