Oliveira Renan F, Vargas Veronica, Plaza-Puche Ana Belén, Alió Jorge L
VISSUM Instituto Oftalmológico de Alicante, Alicante, Spain.
Department of Research, Development and Innovation, VISSUM Instituto Oftalmológico de Alicante, Alicante, Spain.
Eur J Ophthalmol. 2020 Jan;30(1):201-208. doi: 10.1177/1120672118818019. Epub 2018 Dec 17.
To evaluate the long-term objective and subjective results of a trifocal intraocular lens.
Prospective observational study enrolling 24 eyes that underwent uneventful microincisional cataract surgery with bilateral implantation of a diffractive trifocal intraocular lens (FineVision Micro F, PhysIOL, Belgium). After a mean follow-up period of 60 ± 2 months, patients answered the 10-item Near Activity Vision Questionnaire (NAVQ-10) and were assessed for uncorrected and corrected distance, intermediate and near visual acuity, defocus curve, light-distortion analysis, and total ocular aberrometry with a pyramidal wavefront sensor (Osiris, CSO, Italy).
There was significant improvement of uncorrected visual acuity at all distances in all eyes (p < 0.01). The distorsion index and best-fit circle radius were significantly lower in binocular conditions (p = 0.02). Mean Rasch score at NAVQ-10 was 16.29 ± 11.57. Patients reported to be completely, very, and moderately satisfied with their uncorrected near vision in 67%, 25%, and 8% of cases, respectively. We found no significant correlation between patient satisfaction scores at near vision satisfaction questionnaire and other variables such as light-distortion analysis or wavefront aberrometry parameters.
The studied diffractive trifocal intraocular lens provides good and stable long-term results for distance, intermediate, and near vision, with high levels of patient satisfaction at long-term follow-up. The intraocular lens' light-distortion effect is less disturbing in binocular conditions and may have its perception down-modulated by neuroadaptation over time.
评估三焦点人工晶状体的长期客观和主观效果。
前瞻性观察研究,纳入24只眼,这些眼接受了顺利的微切口白内障手术,并双侧植入了衍射三焦点人工晶状体(FineVision Micro F,PhysIOL,比利时)。平均随访60±2个月后,患者回答10项近视力活动问卷(NAVQ-10),并接受未矫正和矫正的远、中、近视力、散焦曲线、光畸变分析以及使用锥形波前传感器(Osiris,CSO,意大利)进行的全眼像差测量。
所有眼在所有距离的未矫正视力均有显著改善(p<0.01)。双眼条件下的畸变指数和最佳拟合圆半径显著更低(p = 0.02)。NAVQ-10的平均Rasch评分为16.29±11.57。分别有67%、25%和8%的患者报告对其未矫正近视力完全、非常和中度满意。我们发现近视力满意度问卷中的患者满意度得分与其他变量(如光畸变分析或波前像差参数)之间无显著相关性。
所研究的衍射三焦点人工晶状体为远、中、近视力提供了良好且稳定的长期效果,长期随访时患者满意度较高。该人工晶状体的光畸变效应在双眼条件下干扰较小,且随着时间推移可能会因神经适应而使其感知下调。