Elbaz Hisham, Werner Liliana, Spiru Bogdan, Sekundo Walter
Ophthalmology Department of the Philipps University of Marburg, Germany.
Intermountain Ocular Research Center, John A. Moran Eye Center - University of Utah, USA.
Saudi J Ophthalmol. 2018 Oct-Dec;32(4):341-345. doi: 10.1016/j.sjopt.2018.03.005. Epub 2018 Mar 21.
We present a case of spontaneous in situ breakage of intracorneal ring segments (ICRSs) 8 years after their implantation in a patient with keratoconus. The patient presented to our clinic with a red and painful right eye that had not improved despite topical steroids and antibiotics. The decision was made to explant the broken ICRSs from the cornea and send them for laboratory analyses, by which manufacturer defects were excluded. It is noteworthy that corneal curvature re-steepening was observed 4 months after ICRS removal despite the performance of crosslinking (CXL) 1 year prior to ICRS implantation.
我们报告了1例圆锥角膜患者,其植入角膜内环片(ICRS)8年后发生了ICRS原位自发破裂。该患者因右眼红肿疼痛前来我院就诊,尽管使用了局部类固醇和抗生素治疗,但症状并未改善。我们决定从角膜上取出破裂的ICRS并送去实验室分析,结果排除了制造商的缺陷。值得注意的是,尽管在植入ICRS前1年已进行了交联(CXL),但在取出ICRS 4个月后仍观察到角膜曲率再次变陡。