Abstract

OBJECTIVES

The HAART ring device has been introduced as a novel strategy to facilitate aortic valve repair. This rigid, elliptical device aims to restore normal leaflet configuration and to provide annular stabilization in the setting of aortic regurgitation. The goal of this preclinical study is to evaluate the in vivo feasibility of 'aortic valve-in-ring' transcatheter aortic valve replacement (TAVR).

METHODS

Six animals {landrace pigs, 87.6 [standard deviation (SD) 4.5] kg} underwent HAART ring implantation (5 cases #19 mm and 1 case #21 mm) via full sternotomy with cardiopulmonary bypass. Seven transfemoral TAVR implantations were performed with the Medtronic EvolutR prosthesis to assess the sizing and outcome (5 cases #23 mm, 1 case #26 mm and 1 case #29 mm).

RESULTS

TAVR implantation was successful in 6 of 7 attempts. Post-dilatation was performed in 1 case without damage of the ring or the valve. One embolization occurred due to oversizing (EvolutR valve 29 mm in HAART ring 19 mm). No clinically relevant postimplantation gradient [7.6 (SD 4.0) mmHg] or regurgitation was detected by invasive and echocardiographic measurements. Postoperative computed tomography scans revealed good device configuration.

CONCLUSIONS

Transcatheter aortic valve-in-ring implantation of a self-expandable TAVR into a rigid aortic annuloplasty ring after aortic valve repair appears feasible. Proper sizing and correct depth of implantation are crucial.

摘要

目的

HAART环装置已作为一种促进主动脉瓣修复的新策略被引入。这种刚性椭圆形装置旨在恢复正常瓣叶形态,并在主动脉瓣反流情况下提供瓣环稳定。本临床前研究的目的是评估“环内主动脉瓣”经导管主动脉瓣置换术(TAVR)在体内的可行性。

方法

6只动物(长白猪,87.6 [标准差(SD)4.5] kg)通过全胸骨切开术并在体外循环下进行HAART环植入(5例#19 mm,1例#21 mm)。使用美敦力EvolutR假体进行7次经股动脉TAVR植入,以评估尺寸选择和结果(5例#23 mm,1例#26 mm,1例#29 mm)。

结果

7次尝试中有6次TAVR植入成功。1例进行了后扩张,未对环或瓣膜造成损伤。1例因尺寸过大发生栓塞(HAART环19 mm中植入EvolutR瓣膜29 mm)。通过有创和超声心动图测量未检测到临床相关的植入后梯度[7.6(SD 4.0)mmHg]或反流。术后计算机断层扫描显示装置形态良好。

结论

在主动脉瓣修复后将可自膨胀TAVR经导管植入刚性主动脉瓣环成形环内似乎是可行的。合适的尺寸选择和正确的植入深度至关重要。