Catheter Cardiovasc Interv. 2019 Nov 15;94(6):806-811. doi: 10.1002/ccd.28160. Epub 2019 Mar 9.
The SAPIEN 3 (S3) transcatheter heart valve (THV) can be over-expanded beyond its labeled diameter. Overexpansion can be achieved with use of either a compliant or non-compliant balloon. Objective data regarding the ability to over-expand older generation balloon expandable valves are limited. We sought to assess the effects of over-expanding the SAPIEN and SAPIEN XT (Edwards Lifesciences, Irvine, CA) valve beyond labeled size (diameter) through an ex-vivo bench study.
We assessed SAPIEN and SAPIEN XT THVs, sized 23/26 mm and 23/26/29 mm, respectively. The SAPIEN THVs were explanted samples. Valves were expanded to nominal dimensions, and then incrementally overexpanded with balloons sized 1-, 2-, and 3-mm larger than the recommended diameter. When an appropriate sized non-compliant balloon was not available, a compliant balloon was utilized. Valves underwent visual and radiographic assessment of overexpansion. SAPIEN THVs with labeled size of 23 and 26 mm could be incrementally overexpanded to midvalve (MV) diameters of 26.7 and 27.4 mm, respectively. SAPIENT XT THVs with labeled size of 23, 26, and 29 mm could be incrementally overexpanded to MV diameters of 26.8, 28.3, and 28.8 mm, respectively. The desired degree of overexpansion was only achieved with use of non-compliant balloons and not with compliant balloons. The outflow of the SAPIEN and SAPIEN XT had larger diameters than the MV and inflow of the THV.
Overexpansion of older generation SAPIEN and SAPIEN XT THVs is possible. Achieving the desired degree of overexpansion was only achieved with use of non-compliant balloons. This has potential implications for the treatment of failed THVs.
SAPIEN 3(S3)经导管心脏瓣膜(THV)可以超过其标定直径进行过度扩张。可以使用顺应性或非顺应性球囊来实现过度扩张。关于过度扩张旧代球囊扩张型瓣膜的能力的客观数据有限。我们试图通过体外研究评估 SAPIEN 和 SAPIEN XT(爱德华兹生命科学公司,加利福尼亚州欧文)瓣膜超过标定尺寸(直径)过度扩张的效果。
我们评估了 SAPIEN 和 SAPIEN XT THV,分别为 23/26mm 和 23/26/29mm。SAPIEN THV 是取出的样本。瓣膜扩张至标称尺寸,然后使用比推荐直径大 1、2 和 3mm 的球囊进行逐步过度扩张。当没有合适尺寸的非顺应性球囊时,使用顺应性球囊。瓣膜进行了过度扩张的视觉和放射评估。标称为 23 和 26mm 的 SAPIEN THV 可以逐步过度扩张至中瓣(MV)直径分别为 26.7 和 27.4mm。标称为 23、26 和 29mm 的 SAPIEN XT THV 可以逐步过度扩张至 MV 直径分别为 26.8、28.3 和 28.8mm。仅使用非顺应性球囊而不是顺应性球囊才能达到所需的过度扩张程度。SAPIEN 和 SAPIEN XT 的流出道直径大于 MV 和 THV 的流入道直径。