在AAI和DDD模式起搏之间切换——其机制是什么?

Switch between AAI and DDD mode pacing-What is the mechanism?

作者信息

Ząbek Andrzej, Małecka Barbara, Nowosielska-Ząbek Ewa, Dębski Maciej, Boczar Krzysztof, Lelakowski Jacek

机构信息

Department of Electrocardiology, The John Paul II Hospital, Kraków, Poland.

Institute of Cardiology, Jagiellonian University Medical College, Kraków, Poland.

出版信息

Ann Noninvasive Electrocardiol. 2019 Jul;24(4):e12648. doi: 10.1111/anec.12648. Epub 2019 Mar 21.

Abstract

The electrocardiogram (ECG) interpretation in patients with implantable pacemaker is often a perplexing problem. The difficulty in the device evaluation increases in the presence of novel timing cycles and additional functions. Authors describe a special function frequently encountered in Medtronic dual-chamber pacemakers and implantable cardioverter-defibrillator devices called managed ventricular pacing (MVP) and demonstrate its performance in the patient with undersensing episodes in ventricular channel. Intermittent ventricular undersensing in the device with MVP feature turned on caused repetitive mode switches between AAI and DDD mode. This report shows unexceptional occurrence of tricky ECG findings in patient with Medtronic dual-chamber device.

摘要

对于植入式起搏器患者的心电图(ECG)解读常常是个令人困惑的问题。在存在新型计时周期和附加功能的情况下,设备评估的难度会增加。作者描述了美敦力双腔起搏器和植入式心脏复律除颤器设备中经常遇到的一种特殊功能,称为心室按需起搏(MVP),并展示了其在心室通道感知不足发作患者中的表现。开启MVP功能的设备中出现间歇性心室感知不足会导致AAI和DDD模式之间反复进行模式切换。本报告显示,美敦力双腔设备患者出现棘手心电图表现的情况并不罕见。

相似文献

1
Switch between AAI and DDD mode pacing-What is the mechanism?
Ann Noninvasive Electrocardiol. 2019 Jul;24(4):e12648. doi: 10.1111/anec.12648. Epub 2019 Mar 21.
4
The reverse mode switch algorithm: how well does it work?
Heart Rhythm. 2013 Aug;10(8):1146-52. doi: 10.1016/j.hrthm.2013.05.025. Epub 2013 May 31.
5
Successful treatment of ventricular fibrillation storm triggered by short-long-short sequence; time to avoid managed ventricular pacing.
J Electrocardiol. 2017 Nov-Dec;50(6):941-943. doi: 10.1016/j.jelectrocard.2017.06.005. Epub 2017 Jun 8.
6
Functional atrial undersensing associated with device algorithm promoting AV conduction.
J Electrocardiol. 2017 Sep-Oct;50(5):696-697. doi: 10.1016/j.jelectrocard.2017.05.009. Epub 2017 May 18.
7
A clinical comparison between a new dual-chamber pacing mode-AAIsafeR and DDD mode.
Am J Med Sci. 2010 Feb;339(2):145-7. doi: 10.1097/MAJ.0b013e3181c3b0c4.
8
A new dual-chamber pacing mode to minimize ventricular pacing.
Pacing Clin Electrophysiol. 2005 Jan;28 Suppl 1:S43-6. doi: 10.1111/j.1540-8159.2005.00095.x.
9
10
Pacing spikes following QRS complexes: What is the mechanism?
Ann Noninvasive Electrocardiol. 2019 Mar;24(2):e12545. doi: 10.1111/anec.12545. Epub 2018 Apr 25.

本文引用的文献

1
Managed ventricular pacing below the lower rate limit during DDD to AAI transition: what is the mechanism?
Pacing Clin Electrophysiol. 2015 Feb;38(2):286-8. doi: 10.1111/pace.12366. Epub 2014 Mar 2.
2
[Atrio-ventricular block in the presence of double-chamber ICD--how is that possible?].
Kardiol Pol. 2013;71(9):969-73. doi: 10.5603/KP.2013.0236.
3
The prescription of minimal ventricular pacing.
Pacing Clin Electrophysiol. 2012 Dec;35(12):1528-36. doi: 10.1111/j.1540-8159.2012.03490.x. Epub 2012 Aug 16.
4
A tale of four atrioventricular intervals.
Europace. 2010 Mar;12(3):441-2. doi: 10.1093/europace/eup343. Epub 2009 Nov 6.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验